Key result
Using a nonlinear function of flow rate for the discharge coefficient reduced the average difference between measured and calculated valve areas from 74% to 3.7% compared to a constant value.
Why the study?
Does modeling the discharge coefficient as a nonlinear function of flow rate improve the accuracy of predicted flow area for a bileaflet mechanical heart valve under simulated stenosis?
Does modeling the discharge coefficient as a nonlinear function of flow rate improve the accuracy of predicted flow area for a bileaflet mechanical heart valve under simulated stenosis?
Modeling the discharge coefficient as a nonlinear function of flow rate significantly improves the accuracy of the Gorlin equation for calculating the area of stenotic mechanical heart valves.
Answers are grounded in this paper and cite their sources.
Should not yet change clinical valve area calculations; leaves open human validation of nonlinear Gorlin modeling.
The possibility that the discharge coefficient (Cd) for a mechanical heart valve (MHV) is affected by valvular stenosis is addressed. A 29 mm bileaflet (Si. Jude Medical) mitral valve is tested on a cardiovascular duplicator (CVD) in its normal state and under two degrees of simulated stenosis. Stenoticity is simulated by bracing the occluders such that full opening is impossible. The pressure drop through the valve is described by a two-term second-order polynomial in flow, from which it is shown that the Cd should be a nonlinear function of the flow rate through the valve. The average difference between measured and calculated areas decreased from 74%, when a constant value of 0.7 was used for the Cd, to 3.7%, when the Cd was a nonlinear function of the flow rate.
No takes yet. Share an insight, caveat, or question.
Souza-Campos et al. (1996) studied Mechanical heart valve stenosis. Nonlinear function of flow rate for discharge coefficient (Cd) vs. Constant value of 0.7 for Cd was evaluated on Average difference between measured and calculated valve areas. Using a nonlinear function of flow rate for the discharge coefficient reduced the average difference between measured and calculated valve areas from 74% to 3.7% compared to a constant value.