Key result
Bidimensional Laser Doppler Anemometry highlighted different flow field features and velocity profiles downstream of four 19 mm sized bileaflet cardiac valves during peak systolic flow.
In vitro analysis of 19 mm bileaflet valve prostheses reveals distinct downstream flow field features among different models at high cardiac output.
Answers are grounded in this paper and cite their sources.
Flow variations among 19 mm bileaflet valves warrant no practice change; leaves open effects on clinical hemodynamics or outcomes.
The investigation of the flow field downstream of a cardiac valve prosthesis is a well established task. In particular turbulence generation is of interest if damage to blood constituents is to be assessed. Several prosthetic valve flow studies are available in literature but they generally concern large-sized prostheses. The FDA draft guidance requires the study of the maximum Reynolds number conditions for a cardiac valve model to assess the worst case in turbulence by choosing both the minimum valve diameter and a high cardiac output value as protocol set up. Within the framework of a national research project regarding the characterization of cardiovascular endoprostheses, the Laboratory of Biomedical Engineering is currently conducting an in-depth study of turbulence generated downstream of bileaflet cardiac valves. Four models of 19 mm sized bileaflet valve prostheses, namely St Jude Medical HP Edwards Tekna, Sorin Bicarbon, and CarboMedics, were studied in aortic position. The prostheses were selected for the nominal annulus diameter reported by the manufacturers without any assessment of the valve sizing method. The hemodynamic function was investigated using a bidimensional LDA system. Results concern velocity profiles during the peak flow systolic phase, at high cardiac output regime, highlighting the different flow field features downstream of the four small-sized cardiac valves.
No takes yet. Share an insight, caveat, or question.
Barbaro et al. (1997) studied Cardiac valve prosthesis flow field (in vitro) (n=4). 19 mm sized bileaflet valve prostheses vs. Comparison among the four valve models was evaluated on Velocity profiles during the peak flow systolic phase at high cardiac output regime. Bidimensional Laser Doppler Anemometry highlighted different flow field features and velocity profiles downstream of four 19 mm sized bileaflet cardiac valves during peak systolic flow.
Synapse has enriched 10 related papers. Consider them for comparative context: