Key result
Row-column array ultrasound accurately measures 3D full-thickness myocardial strain in porcine tissue with ~2% error.
Why the study?
To implement and validate a 3D volume imaging sequence and 3D strain estimation procedure for enhanced biaxial mechanical testing of excised ventricular myocardium.
Does 3D volume ultrasound imaging accurately measure tissue strain compared to optical camera imaging in excised ventricular porcine myocardium?
Does 3D volume ultrasound imaging accurately measure tissue strain compared to optical camera imaging in excised ventricular porcine myocardium?
3D full-thickness strain measurement with ultrasound imaging accurately measures tissue deformation in excised porcine myocardium, offering enhanced biomechanical insights over conventional surface imaging.
Answers are grounded in this paper and cite their sources.
Validates 3D ultrasound strain feasibility ex vivo; leaves open in vivo validation and clinical translation.
OBJECTIVE: To implement and validate a 3D volume imaging sequence and 3D strain estimation procedure for enhanced biaxial mechanical testing of excised ventricular myocardium. METHODS: One specimen of right and one of left ventricular excised porcine myocardium were tested using dual-loading protocol quasistatic biaxial mechanical testing. During biaxial testing, volume ultrasound (US) images were acquired using a row-column addressed array probe using a synthetic aperture imaging sequence. Volume US images were used to compute tissue deformation using 3D correlation-based US speckle tracking. US-derived tissue strains were validated against repeated measurements using conventional optical camera imaging of the specimen surface deformation. RESULTS: Speckle tracking yielded high-fidelity maps of tissue deformation in 3D throughout the entire sample volume. US-derived tissue strain is in good agreement with ground-truth camera-derived surface strain measurements (root mean square error is 1.6% strain). CONCLUSION: The 3D full-thickness strain measurement with US imaging is accurate and can enhance biomechanical insights from biaxial experimentation, especially in large tissues such as porcine and human myocardium where assumptions of plane stress and incompressibility may not apply.
No takes yet. Share an insight, caveat, or question.
Navy et al. (2025) studied Excised ventricular porcine myocardium (n=2). 3D volume ultrasound imaging and 3D strain estimation using a row-column addressed array probe vs. Conventional optical camera imaging of specimen surface deformation was evaluated on Tissue deformation/strain measurement accuracy (root mean square error). 3D full-thickness strain measurement using ultrasound imaging with a row-column array accurately estimated tissue deformation in excised porcine myocardium with a root mean square error of 1.6% strain.