Key result
Row-column array ultrasound accurately estimates 3D tissue strain vs optical imaging with ~1.6% error.
Why the study?
To implement and validate a 3D volume imaging sequence and 3D strain estimation procedure for enhanced biaxial mechanical testing of excised ventricular myocardium.
Does 3D volume ultrasound imaging accurately measure tissue strain compared to standard optical camera imaging in excised porcine myocardium?
Does 3D volume ultrasound imaging accurately measure tissue strain compared to standard optical camera imaging in excised porcine myocardium?
Answers are grounded in this paper and cite their sources.
High-fidelity 3D strain measurement with ultrasound imaging accurately estimates tissue deformation in excised porcine myocardium, enhancing biomechanical insights from biaxial experimentation.
Objective To implement and validate a 3D volume imaging sequence and 3D strain estimation procedure for enhanced biaxial mechanical testing of excised ventricular myocardium. Methods One right- and one left-ventricular excised porcine myocardium specimens were tested using dual-loading protocol quasi-static biaxial mechanical testing. During biaxial testing, volume ultrasound images were acquired using a row-column addressed array probe using a synthetic aperture imaging sequence. Volume ultrasound images were used to compute tissue deformation using 3D correlation-based ultrasound speckle tracking. Ultrasound-derived tissue strains were validated against repeated measurements using typical standard optical camera imaging of the surface deformation of the specimen. Results Synthetic aperture imaging with the RCA yielded artifact-free and uniform ultrasound speckle for high-fidely speckle tracking of tissue deformation. Ultrasound-derived tissue strain is in good agreement with ground-truth camera-derived surface strain measurements (root-mean-square error is 1.6% strain). Conclusion High-fidelity 3D strain measurement with ultrasound imaging is accurate and can enhance biomechanical insights from biaxial experimentation, especially in large tissues such as porcine and human myocardium where assumptions of plane stress and incompressibility may not apply.
No takes yet. Share an insight, caveat, or question.
Navy et al. (2024) studied Excised ventricular myocardium (n=2). 3D tissue strain measurement using a row-column array ultrasound vs. Standard optical camera imaging was evaluated on Tissue strain measurement accuracy (root-mean-square error). Ultrasound-derived 3D tissue strain measurement using a row-column array accurately estimated tissue deformation compared to standard optical camera imaging.