Key result
Dual-chamber para-His pacing acutely preserved LV systolic performance and mechanical synchrony compared with right ventricular septal pacing.
Why the study?
Does dual-chamber para-His pacing preserve left ventricular performance and mechanical synchrony compared to right ventricular septal pacing in patients with high-grade atrioventricular block?
Does dual-chamber para-His pacing preserve left ventricular performance and mechanical synchrony compared to right ventricular septal pacing in patients with high-grade atrioventricular block?
In patients with high-grade AV block, acute para-His pacing preserves left ventricular systolic performance and mechanical synchrony better than right ventricular septal pacing.
Answers are grounded in this paper and cite their sources.
Acute LV benefit in cohort supports further study; leaves open long-term outcomes and practice change pending RCTs.
AIM: To compare changes in left ventricular (LV) performance and mechanical synchrony between atrial-based pacing with intrinsic conduction (AAI), dual-chamber para-His Pacing (DDD-PHP) and dual-chamber right ventricular septal pacing (DDD-RVS) in patients with high-grade atrioventricular block (AVB). METHODS AND RESULTS: Patients with high-grade AVB and QRS <120 ms, who had temporary intrinsic atrioventricular (AV) conduction the day after the implantation were included in the study. All patients received a biventricular pacemaker with a para-His lead in the LV port, and a RVS lead in the right ventricular port. Left ventricular three-dimensional echocardiograms, LV outlet tract-velocity time integrals (LVOT-VTI), and LV synchrony with tissue Doppler imaging were recorded during AAI, DDD-PHP, and DDD-RVS. Eleven patients were included. The mean LVOT-VTI was significant lower during DDD-RVS (19.2 ± 5.5 cm) as compared with DDD-PHP (21.4 ± 5.5 cm), P = 0.006 and AAI (21.6 ± 6.8 cm), P = 0.016. The LVEF was higher during AAI than during DDD-PHP (P= 0.02) and DDD-RVS (P< 0.01). The maximal time to peak velocity between basal segments was significant longer with DDD-RVS (95 ± 26 ms) than with AAI (72 ± 30 ms), P = 0.028, whereas no difference was observed between AAI and DDD-PHP (81 ± 42 ms), P = 0.20. CONCLUSIONS: Acutely, DDD-PHP preserves LV systolic performance and mechanical synchrony as compared with DDD-RVS.
No takes yet. Share an insight, caveat, or question.
Kronborg et al. (2011) studied High-grade atrioventricular block (n=11). Dual-chamber para-His Pacing (DDD-PHP) vs. Dual-chamber right ventricular septal pacing (DDD-RVS) and atrial-based pacing (AAI) was evaluated on Mean LVOT-VTI (p=0.006). Dual-chamber para-His pacing acutely preserved LV systolic performance and mechanical synchrony compared with right ventricular septal pacing.
Synapse has enriched 10 related papers. Consider them for comparative context: