Key result
Non-invasive iECG shows moderate ~0.54 correlation with invasive epicardial mapping alongside considerable absolute timing differences.
Why the study?
To quantitatively compare a novel non-invasive equivalent dipole layer based inverse electrocardiography approach with invasive electro-anatomical mapping during sinus rhythm to enable functional substrate imaging and sudden cardiac death risk stratification.
Does non-invasive inverse electrocardiography accurately estimate ventricular activation sequences compared to invasive electro-anatomical mapping in patients referred for EAM?
Observational (n=13)
Does non-invasive inverse electrocardiography accurately estimate ventricular activation sequences compared to invasive electro-anatomical mapping in patients referred for EAM?
This first-in-human quantitative comparison shows that non-invasive inverse electrocardiography has moderate correlation with invasive electro-anatomical mapping for ventricular activation sequences, but requires further refinement for clinical use.
Answers are grounded in this paper and cite their sources.
Noninvasive inverse ECG requires further refinement before clinical use; leaves open its role in ventricular activation mapping.
This study presents a novel non-invasive equivalent dipole layer (EDL) based inverse electrocardiography ( i ECG) technique which estimates both endocardial and epicardial ventricular activation sequences. We aimed to quantitatively compare our i ECG approach with invasive electro-anatomical mapping (EAM) during sinus rhythm with the objective of enabling functional substrate imaging and sudden cardiac death risk stratification in patients with cardiomyopathy. Thirteen patients (77% males, 48 ± 20 years old) referred for endocardial and epicardial EAM underwent 67-electrode body surface potential mapping and CT imaging. The EDL-based i ECG approach was improved by mimicking the effects of the His-Purkinje system on ventricular activation. EAM local activation timing (LAT) maps were compared with i ECG-LAT maps using absolute differences and Pearson’s correlation coefficient, reported as mean ± standard deviation [95% confidence interval]. The correlation coefficient between i ECG-LAT maps and EAM was 0.54 ± 0.19 [0.49–0.59] for epicardial activation, 0.50 ± 0.27 [0.41–0.58] for right ventricular endocardial activation and 0.44 ± 0.29 [0.32–0.56] for left ventricular endocardial activation. The absolute difference in timing between i ECG maps and EAM was 17.4 ± 7.2 ms for epicardial maps, 19.5 ± 7.7 ms for right ventricular endocardial maps, 27.9 ± 8.7 ms for left ventricular endocardial maps. The absolute distance between right ventricular endocardial breakthrough sites was 30 ± 16 mm and 31 ± 17 mm for the left ventricle. The absolute distance for latest epicardial activation was median 12.8 [IQR: 2.9–29.3] mm. This first in-human quantitative comparison of i ECG and invasive LAT-maps on both the endocardial and epicardial surface during sinus rhythm showed improved agreement, although with considerable absolute difference and moderate correlation coefficient. Non-invasive i ECG requires further refinements to facilitate clinical implementation and risk stratification.
No takes yet. Share an insight, caveat, or question.
Roudijk et al. (2021) conducted an observational in Cardiomyopathy (n=13). Non-invasive equivalent dipole layer (EDL) based inverse electrocardiography (iECG) vs. Invasive electro-anatomical mapping (EAM) was evaluated on Correlation coefficient and absolute differences in local activation timing (LAT) between iECG and EAM (r=0.54 (epicardial), 95% CI 0.49-0.59). Non-invasive inverse electrocardiography showed moderate correlation with invasive mapping for epicardial activation alongside considerable absolute timing differences.
Synapse has enriched 10 related papers. Consider them for comparative context: