Why the study?
Can a novel calculation method based on radial applanation tonometry accurately estimate central aortic pressure?
Can a novel calculation method based on radial applanation tonometry accurately estimate central aortic pressure?
A novel computational method using radial applanation tonometry can accurately estimate central aortic pressure, providing a non-invasive alternative to direct measurement.
Answers are grounded in this paper and cite their sources.
May support non-invasive aortic pressure estimation; leaves open clinical utility pending larger validation studies.
One-dimensional arterial flow simulations are suitable to estimate the aortic pressure from peripheral measurements in a patient-specific arterial network. This study introduces a reduction of the system parameters, and a novel calculation method to estimate the patient-specific set and the aortic curve based on radial applanation tonometry. Peripheral and aortic pressure curves were measured in patients, optimization were carried out. The aortic pressure curves were reproduced well, with an overestimation of the measured Systolic and Mean Blood Pressures on average by 0.6 and 0.5 mmHg respectively, and the Root Mean Square Difference of the curves was 3 mmHg on average.
No takes yet. Share an insight, caveat, or question.
Gyürki et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: