Key result
TAVR maintains ~94% 5-year freedom from SVD, with age and valve size predicting risk.
Why the study?
As TAVR expands to younger and lower-risk patients, long-term durability is crucial, prompting evaluation of the incidence and risk factors of structural valve degeneration at midterm follow-up.
Cohort (n=344)
No
In a veteran population undergoing TAVR, structural valve degeneration occurred at a low but detectable rate at 5 years, with younger age and smaller valve size identified as independent risk factors.
May inform TAVR valve selection in younger veterans; leaves open need for longer-term validation in broader cohorts.
OBJECTIVE: Transcatheter aortic valve replacement (TAVR), previously reserved for patients of intermediate to prohibitive surgical risk, has now been expanded to patients of any surgical risk with severe aortic stenosis. Bioprostheses are prone to structural valve degeneration (SVD), a progressive and multifactorial process that limits valve durability. As the population undergoing TAVR shifts toward a lower-risk and younger profile, long-term durability is a crucial determinant for patient outcomes. Our objective was to determine the incidence and risk factors of SVD at midterm follow-up in a veteran TAVR population. METHODS: Patients undergoing TAVR at our federal facility were retrospectively evaluated for SVD and other endpoints with standardized consensus criteria. Multivariable Cox proportional hazards analysis was performed to evaluate risk factors for mortality and SVD. RESULTS: From 2013 to 2020, 344 patients (median age, 78 years) underwent TAVR. Survival from all-cause mortality was 91.3% at 1 year, 75.1% at 3 years, and 61.7% at 5 years. Cumulative freedom from SVD was 98.2% at 1 year, 96.5% at 3 years, and 93.7% at 5 years. All 13 patients with SVD met hemodynamic criteria, and 1 required intervention. Median time to hemodynamic SVD was 1.04 years. Independent risk factors for SVD included age (hazard ratio [HR] = 0.92, 95% confidence interval [CI]: 0.86 to 0.99) and valve size (HR = 0.19, 95% CI: 0.04 to 0.89). CONCLUSIONS: SVD was evident at a low but detectable rate at 5-year follow-up. Further understanding of TAVR biomechanics as well as continued longer-term follow-up will be essential for informing patient-specific risk of SVD.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2022) conducted a cohort in Severe aortic stenosis (n=344). Transcatheter aortic valve replacement (TAVR) was evaluated on Cumulative freedom from structural valve degeneration at 5 years (95% CI 89.0-98.3). In veterans undergoing transcatheter aortic valve replacement, cumulative freedom from structural valve degeneration was 93.7% at 5 years, with age and valve size identified as independent risk factors.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: