Key result
Bipolar atrial electrogram amplitude during AF ablation shows beat-to-beat variability up to ~1.74 mV.
Why the study?
Variability of bipolar atrial electrogram amplitude may affect voltage maps created during ablation procedures and the extent of ablations.
Observational (n=11)
Bipolar atrial electrogram amplitude recorded during AF ablation exhibits clinically relevant beat-to-beat variability, which is greater during remapping.
Variability warrants caution against single-beat voltage reliance in AF ablation; leaves open optimal multi-beat protocols for research.
INTRODUCTION: Variability of the bipolar atrial electrogram amplitude may affect voltage maps created during ablation procedures, and thus also the extent of ablations. Therefore, the aim of the study was to assess the beat-to-beat electrogram amplitude variability in the left atrium in patients undergoing atrial fibrillation ablation. METHODS: In 11 patients undergoing ablation for atrial fibrillation, 362 mapping points were collected in two series. At each point, three consecutive beats were recorded and verified including the bipolar electrogram amplitude, contact force (CF), and orientation of the catheter tip. The repeatability and reproducibility of obtained measurements between consecutive beats and series were assessed by the Pearson correlation coefficient (r), the Bland-Altman test, repeatability coefficient (RC), relative standard deviation (RSD), and concordance correlation coefficient (CCC). RESULTS: A total of 1086 beats were analyzed. The correlation coefficient for bipolar atrial electrogram amplitude for the first two beats, and for the first and the third beats were 0.94 and 0.86, respectively. The average of differences between the first two beats and between the first and the third beats were 0.06 and 0.13 mV with 95% limits of agreement (LoA) within ±0.98 and ±1.74 mV, respectively. For CF values ≤5 and ≥20 g, the 95% LoA were narrower compared to other CF ranges and were ±0.49 and ±0.71 mV from the average value, respectively. When the analyzes were performed within the predefined ranges of bipolar electrogram amplitude: 0.05-1; 1-2; 2-3 mV, the 95% LoA were within ±0.33, ±0.98, and ±0.84 mV from the average value, respectively. RC and RSD were 1.41 mV and 20.8%, respectively. For repeated measurement between series, CCC ranged from 0.67 to 0.71 and the 95% LoA were within ±2.7 to 2.9 mV from the average value. CONCLUSION: Bipolar atrial electrogram amplitude recorded at a given site during ablation procedures is variable to an extent that may be clinically relevant. The magnitude of the observed variability is greater during remapping.
No takes yet. Share an insight, caveat, or question.
Hangiel et al. (2022) conducted an observational in Atrial fibrillation (n=11). Atrial electrogram mapping was evaluated on Beat-to-beat electrogram amplitude variability. Bipolar atrial electrogram amplitude recorded during atrial fibrillation ablation exhibits clinically relevant beat-to-beat variability, with 95% limits of agreement up to ±1.74 mV.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: