Key result
Single-lead DX ICD bipolar atrial pacing shows no significant capture benefit over unipolar configurations.
Why the study?
While atrial sensing performance of the single-lead DX ICD with an integrated atrial floating dipole has been extensively documented, the ability of the floating atrial dipole to achieve atrial capture had not previously been investigated.
Does bipolar pacing through an integrated atrial floating dipole achieve superior acute atrial capture and lower pacing thresholds compared to unipolar configurations in patients undergoing single-lead DX ICD implantation or replacement?
Observational (n=25)
Yes
Does bipolar pacing through an integrated atrial floating dipole achieve superior acute atrial capture and lower pacing thresholds compared to unipolar configurations in patients undergoing single-lead DX ICD implantation or replacement?
This study provides proof-of-concept evidence that acute atrial capture via the floating atrial dipole of a single-lead DX ICD is feasible in 88% of patients using a bipolar pacing configuration, supporting future development of single-lead dual-chamber pacing.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for bipolar pacing in DX ICDs; leaves open prospective validation before clinical adoption.
INTRODUCTION: The single-lead DX implantable cardioverter-defibrillator (ICD) combines the advantages of a conventional single-lead ICD with atrial sensing capability provided by an integrated atrial floating dipole. While atrial sensing performance has been extensively documented, the ability of the floating atrial dipole to achieve atrial capture has not previously been investigated. METHODS AND RESULTS: Consecutive patients undergoing de novo implantation or generator replacement of a single-lead DX ICD system at two Italian centers were prospectively enrolled. Acute atrial pacing was assessed using three configurations: bipolar pacing and two unipolar configurations employing either the proximal or distal atrial electrode. Atrial capture, pacing threshold (PT), impedance, and phrenic nerve stimulation were evaluated at pulse widths of 0.4 and 1.0 ms. Twenty-five patients (92% male; mean age 66.2 ± 10.1 years) were included. Acute atrial capture was achieved in 22 patients (88%) with bipolar pacing, compared with 18 (72%) and 17 (68%) using proximal and distal unipolar pacing, respectively (p = 0.122). Bipolar pacing yielded significantly lower PTs than unipolar configurations at both pulse widths (p < 0.020). Median atrial sensing amplitude measured after connection to the ICD generator was 3.9 mV (2.8-5.1). CONCLUSION: Atrial capture through the floating atrial dipole of a single-lead DX ICD system is feasible in the majority of patients, with bipolar pacing providing the highest capture rate and the lowest PTs. These findings provide proof-of-concept evidence supporting further development of single-lead dual-chamber sensing and pacing strategies using the DX platform.
No takes yet. Share an insight, caveat, or question.
Cerini et al. (2026) conducted an observational in Patients undergoing de novo implantation or generator replacement of a single-lead DX ICD system (n=25). Bipolar atrial pacing via floating atrial dipole of a single-lead DX ICD vs. Proximal and distal unipolar atrial pacing configurations was evaluated on Acute atrial capture (p=0.122). Bipolar pacing via the single-lead DX ICD floating atrial dipole achieved acute atrial capture in 88% of patients vs 72% and 68% with unipolar pacing, with lower pacing thresholds.