Key result
Biotronik DX single-lead ICD provides reliable atrial sensing without the risks of an atrial lead.
Why the study?
The optimal device strategy between dual- versus single-chamber ICDs in patients without conventional pacing indications remains debatable, and the benefit of prophylactic pacing is undocumented.
Answers are grounded in this paper and cite their sources.
Demonstrates reliable atrial sensing via single-lead defibrillators with floating dipoles, indicating dual-chamber diagnostic benefits without additional lead complications.
In patients known to be a high risk for sudden cardiac arrest, implantable cardioverter defibrillators (ICD) are a proven therapy to reduce risk of death. However, in patients without conventional indications for pacing, the optimal strategy for type of device, dual- versus single-chamber, remains debatable. The benefit of prophylactic pacing in this category of patients has never been documented. Although available atrial electrograms in a dual chamber system improve interpretation of stored arrhythmia events, allow monitoring of atrial fibrillation and may potentially reduce the risk of inappropriate shocks by enhancing automated arrhythmia discrimination, the use of dual-chamber ICDs has a number of disadvantages. The addition of an atrial lead adds complexity to implantation and extraction procedures, increases procedural cost and is associated with a higher risk of periprocedural complications. The single lead pacing system with ability to sense atrial signals via floating atrial electrodes (VDD) clinically became available in early 1980's but did not gain much popularity due to inconsistent atrial sensing and concerns about the potential need for an atrial lead if sinus node fails. Most ICD patients do not have indications for pacing at implantation and subsequent risk of symptomatic bradycardia seems to be low. The concept of atrial sensing via floating electrodes has recently been revitalized in the Biotronik DX ICD system (Biotronik, SE & Co., Berlin, Germany) aiming to provide all of the potential advantages of available atrial electrograms without the risks and incremental cost of an additional atrial lead. Compared to a traditional VDD pacing system, the DX ICD system uses an optimized (15 mm) atrial dipole spacing and improved atrial signal processing to offer more reliable atrial sensing. The initial experience with the DX system indicates that the clinically useful atrial signal amplitude in sinus rhythm remains stable over time. Future studies are needed to determine reliability of atrial sensing during tachyarrhythmias, particularly atrial fibrillation as well as clinical utility and cost-effectiveness of this technology in different populations of patients.
No takes yet. Share an insight, caveat, or question.
WORDEN et al. (2016) conducted a review in High risk for sudden cardiac arrest without conventional indications for pacing. Single-lead implantable cardioverter-defibrillator (ICD) system with a floating atrial dipole (Biotronik DX) vs. Dual-chamber ICDs and traditional VDD pacing systems was evaluated. The single-lead Biotronik DX ICD system uses an optimized 15 mm atrial dipole spacing to provide reliable atrial sensing without the risks and incremental costs of an additional atrial lead.
Synapse has enriched 10 related papers. Consider them for comparative context: