Key result
Novel automated template-matching algorithm identifies simulated PVCs with ~99% specificity and high pace-mapping resolution.
Why the study?
Does a novel automated template-matching algorithm accurately recognize and spatially discriminate simulated PVCs in a porcine model?
Does a novel automated template-matching algorithm accurately recognize and spatially discriminate simulated PVCs in a porcine model?
Effect estimate: c-statistic 0.93
p-value: p=<0.001
A novel automated template-matching algorithm demonstrated high accuracy and a 2 mm spatial resolution for pace-mapping in a porcine model, indicating potential to assist in PVC ablation.
Supports automated PVC localization tools in preclinical models; leaves open human validation and clinical adoption.
INTRODUCTION: Template-matching algorithms are routinely used in the catheter ablation of patients with premature ventricular contractions (PVCs). However, systematic analysis of the accuracy and spatial resolution of such systems is lacking. Therefore, the aim of this evaluation was to perform a systematic in vivo validation of performance of a novel automated template-matching algorithm. METHODS AND RESULTS: In a porcine model, paced beats simulating PVCs from different origins were investigated. The ability to discriminate between sinus rhythm and PVCs was tested by simulating PVCs using sequential pacing from different cardiac chambers. The accuracy of the algorithm in correctly classifying PVCs was reviewed by an independent investigator. In addition, the spatial resolution of pace matching was evaluated by assessing the QRS morphology discrimination at a distance of 0, 2, 4, and 6 mm of a simulated PVCs focus. The specificity of the algorithm for recognizing simulated PVCs was 99.6% and the sensitivity was 85.3%. There was a significant difference in the discrimination metric discrimination metric (with 0% being a perfect match and 100% being no correlation) between PVC origin (median 0%, interquartile range (IQR) 0-2%) versus at 2 mm (5%, IQR 2-7%), 4 mm (16%, IQR 11-21%), and 6 mm (24%, IQR 19-28%, P < 0.001 for all). The c-statistic for discrimination between PVC origin and a distance ≥ 2 mm was 0.93. CONCLUSIONS: Automated template matching had high specificity and sensitivity, with good spatial discrimination and a pace-mapping resolution in range of 2 mm. Clinical application of this algorithm may assist in the interventional treatment of patients with PVCs.
No takes yet. Share an insight, caveat, or question.
Kosiuk et al. (2017) studied Premature ventricular contractions. Automated template-matching algorithm vs. Sinus rhythm and varying distances from PVC focus was evaluated on Specificity and sensitivity for recognizing simulated PVCs and spatial resolution of pace matching (c-statistic 0.93, p=<0.001). A novel automated template-matching algorithm demonstrated 99.6% specificity and 85.3% sensitivity for recognizing simulated PVCs, with a pace-mapping resolution of 2 mm (c-statistic 0.93).