Key result
LAT-hybrid module fails to reduce PVC offset versus conventional mapping despite correcting positional shifts.
Why the study?
Displacement in catheter position between sinus rhythm and premature ventricular contractions complicates 3D mapping localization, creating a need for an automated algorithm to annotate local activation time sites at corresponding sinus rhythm sites.
Does an automated mapping algorithm (LAT-hybrid module) improve procedural efficiency and accuracy compared to conventional mapping in patients undergoing PVC ablation?
Cohort (n=38)
Unblinded
No
Does an automated mapping algorithm (LAT-hybrid module) improve procedural efficiency and accuracy compared to conventional mapping in patients undergoing PVC ablation?
Absolute Event Rate: 6.96% vs 9.48%
p-value: p=0.1
An automated PVC mapping algorithm (LAT-hybrid) accurately corrects for positional shifts and improves procedural efficiency by reducing RF time.
Automated PVC mapping algorithm matches manual accuracy with shorter RF times; leaves open prospective validation before routine use.
Background Catheter mapping and ablation of premature ventricular contractions (PVCs) requires accurate annotation of earliest local activation time (LAT), but displacement in catheter position between sinus rhythm (SR) and the PVC complicates three-dimensional mapping localization. An automated algorithm to annotate LAT sites at the corresponding sinus rhythm sites would provide an alternative to manual annotation and improve procedural efficiency. Methods A retrospective single center study assessed 64 patients undergoing catheter ablation of PVCs. We divided the study patients into two groups: the patients underwent RF ablation using the CARTO 3 version 7 with LAT-hybrid™ module (hybrid group) and the other patients using the CARTO 3 version 6 without LAT-hybrid™ module (conventional group). Results The primary results of this study demonstrate that the mean hybrid distance is 4.03 ± 2.33 mm, and this automated algorithm can correct for the positional shift with accuracy comparable to manual correction. Moreover, this algorithm showed significantly shorter RF time and the shorter PVC offset compared to the conventional method. Conclusion PVC mapping with the LAT-hybrid™ module demonstrated that an automated algorithm can map in more detail with no compromise in accuracy. The LAT-hybrid™ module showed significantly shorter RF time and PVC offset from the earliest LAT in LVOT cases. Graphical Abstract Spatial displacement of earliest site of activation for premature ventricular contractions (PVCs) relative to sinus rhythm beats in high-risk para-His bundle region. Automated spatial correction (LAT-Hybrid) provides a more focal area of interest 1.4mm posterior to the His region adjacent to the tricuspid annulus. The LAT-Hybrid automated algorithm directed ablation to the successful site with single burn elimination.
No takes yet. Share an insight, caveat, or question.
Ishida et al. (2025) conducted a cohort in Premature ventricular contractions (n=38). LAT-hybrid module (CARTO 3 version 7) vs. CARTO 3 version 6 without LAT-hybrid module (conventional method) was evaluated on PVC offset from the earliest local activation time (mm) (p=0.1). The LAT-hybrid module automatically corrected positional shifts with a mean hybrid distance of 4.03 mm, but did not significantly reduce overall PVC offset from the earliest local activation time (6.96 vs 9.48 mm, p=0.1) compared to conventional mapping.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: