Key result
Modified omnipolar mapping exceeds 85% fibrosis detection accuracy versus standard maps in simulated atrial tissue.
Why the study?
A modified omnipolar electrogram method was proposed to reduce orientation-dependent residual components found in the standard omnipolar approach for substrate and conduction velocity mapping.
A modified omnipolar electrogram mapping strategy improves fibrosis detection and robustness against noise in simulated 2D tissue compared to standard omnipolar and bipolar mapping.
Answers are grounded in this paper and cite their sources.
Should not yet alter clinical mapping; hypothesis-generating for fibrosis detection in simulated atrial models.
Introduction: The omnipolar electrogram method was recently proposed to try to generate orientation-independent electrograms. It estimates the electric field from the bipolar electrograms of a clique, under the assumption of locally plane and homogeneous propagation. The local electric field evolution over time describes a loop trajectory from which omnipolar signals in the propagation direction, substrate and propagation features, are derived. In this work, we propose substrate and conduction velocity mapping modalities based on a modified version of the omnipolar electrogram method, which aims to reduce orientation-dependent residual components in the standard approach. Methods: A simulated electrical propagation in 2D, with a tissue including a circular patch of diffuse fibrosis, was used for validation. Unipolar electrograms were calculated in a multi-electrode array, also deriving bipolar electrograms along the two main directions of the grid. Simulated bipolar electrograms were also contaminated with real noise, to assess the robustness of the mapping strategies against noise. The performance of the maps in identifying fibrosis and in reproducing unipolar reference voltage maps was evaluated. Bipolar voltage maps were also considered for performance comparison. Results: Results show that the modified omnipolar mapping strategies are more accurate and robust against noise than bipolar and standard omnipolar maps in fibrosis detection (accuracies higher than 85 vs. 80% and 70%, respectively). They present better correlation with unipolar reference voltage maps than bipolar and original omnipolar maps (Pearson's correlations higher than 0.75 vs. 0.60 and 0.70, respectively). Conclusion: The modified omnipolar method improves fibrosis detection, characterization of substrate and propagation, also reducing the residual sensitivity to directionality over the standard approach and improving robustness against noise. Nevertheless, studies with real electrograms will elucidate its impact in catheter ablation interventions.
No takes yet. Share an insight, caveat, or question.
Riccio et al. (2021) studied Atrial fibrillation and atrial fibrosis. Modified omnipolar electrogram (MOP-EGM) mapping strategy vs. Bipolar and standard omnipolar electrogram mapping was evaluated on Accuracy in fibrosis detection. The modified omnipolar mapping strategies achieved higher accuracy in fibrosis detection compared to bipolar and standard omnipolar maps in a simulated 2D atrial tissue model.