September 1, 2005Pacing and Clinical Electrophysiology276 citations

Complications of Implantable Cardioverter Defibrillator Therapy in 440 Consecutive Patients

View Full Paper
PAPeter AlterSWS WaldhansEPEVELINE PLACHTA

Structured PICO

What is the incidence of complications associated with implantable cardioverter defibrillator therapy in patients undergoing first implantation?

P
Population
440 consecutive patients who underwent first implantation of an implantable cardioverter defibrillator (ICD) system for primary or secondary prevention of sudden cardiac death
I
Intervention
Implantable cardioverter defibrillator (ICD) therapy (pectoral nonthoracotomy ICD lead systems in all but one patient)
O
Outcome
ICD-related complicationssafety

More than one quarter of patients receiving an ICD experience complications over a mean follow-up of almost 4 years, though serious complications like death or infection are rare.

Abstract

Background: Although more than 150,000 implantable cardioverter defibrillators (ICDs) are implanted yearly worldwide, only few studies systematically examined complications of ICD therapy in large patient cohorts. Methods: We prospectively analyzed ICD‐related complications in 440 consecutive patients who underwent first implantation of an ICD system for primary or secondary prevention of sudden cardiac death within the last 10 years at our institution. All study patients received pectoral nonthoracotomy ICD lead systems with the exception of one patient who had an artificial tricuspid valve. Results: During 46 ± 37 months follow‐up, 136 of 440 patients (31%) experienced at least one complication including implant procedure‐related complications in 43 patients (10%), ICD generator‐related complications in 28 patients (6%), lead‐related complications in 52 patients (12%), and inappropriate shocks in 54 patients (12%). The most serious complications included one perioperative death due to heart failure (0.2%), two ICD system infections necessitating device removal (0.5%) and two perioperative cerebrovascular strokes (0.5%). Conclusions: We conclude that more than one quarter of ICD patients experience complications during a mean follow‐up of almost 4 years, although serious complications such as intraoperative death or ICD system infections are rare in patients with nonthoracotomy ICD systems. Recognition of these complications is the prerequisite for advances in ICD technology and management strategies to avoid their recurrence.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Alter et al. (2005) studied this question.

synapsesocial.com/papers/6a0f111cb9cfc04f9247bb1ehttps://doi.org/10.1111/j.1540-8159.2005.00195.x
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Complications of Implantable Cardioverter Defibrillator Therapy: Follow‐Up of 241 Patients1993 · 97 citations
  2. 2Complications of Third‐Generation Implantable Cardioverter Defibrillator Therapy1999 · 76 citations
  3. 3Evaluating current implantable cardioverter defibrillator implantation procedures: can common complications be minimised?2016
  4. 4Low Incidence of Lead Related Complications Associated with Nonthoracotomy Implantable Cardioverter Defibrillator Systems1995 · 64 citations
  5. 5Long-term complications, reoperations and survival following cardioverter-defibrillator implant2017 · 45 citations