Nonthoracotomy ICD lead systems are associated with a low complication rate (8%), though complications often require reoperation and may be asymptomatic.
Answers are grounded in this paper and cite their sources.
Observational data support nonthoracotomy ICD lead safety with monitoring; leaves open need for prospective trials on reoperation and asymptomatic detection.
UNLABELLED: Implantable cardioverter defibrillators (ICDs) are increasingly being implanted without the need for thoracotomy. Long-term lead performance and stability were evaluated in 150 consecutive patients in whom 1 of 3 nonthoracotomy ICD lead systems was implanted over a 3-year period from September 1990. RESULTS: Twelve (8%) patients (7 males, 5 females) experienced 13 lead complications during a follow-up period of 12 +/- 10 months. Complications were related to intracardiac leads in 7 (4 dislodgments, 2 fractures, 1 right ventricular perforation) and patch leads in 6 (2 folding, 1 fracture, 1 erosion, and 2 hematomas) cases. Freedom from lead related complications at 1 year was 92% (95% confidence interval, 86%-95%). A significant difference in freedom from lead complications between the two most frequently implanted lead systems was observed (P = 0.02). Complication rates were similar in the initial 75 and the more recent 75 implants (P = 0.5). The median time between lead implant and detection of complications was 37 days (range 3-1,147). Complications were diagnosed before hospital discharge in only two cases. In five patients, complications were asymptomatic and in three of these, reoperation was required due to inadequate defibrillation thresholds. Reoperation was necessary in 9 of 12 patients. CONCLUSIONS: Nonthoracotomy ICD lead systems are associated with a low complication rate. Complications may or may not cause symptoms, usually occur after hospital discharge, and require reoperation. Complications are not related to a "learning curve." There is a significant difference in performance between different lead systems.
No takes yet. Share an insight, caveat, or question.
Fahy et al. (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: