Key result
Pectoral implantation of large ICDs with transvenous leads resulted in a 13% overall incidence of insulation erosion in one lead system at 50 months of follow-up.
Why the study?
What is the incidence of late lead complications in patients with transvenous defibrillator leads and ICDs implanted in the pectoral region?
Cohort (n=132)
What is the incidence of late lead complications in patients with transvenous defibrillator leads and ICDs implanted in the pectoral region?
Pectoral implantation of large ICD generators with long transvenous leads is associated with a moderate risk of late insulation breakdown due to pressure against the leads in the pocket.
Answers are grounded in this paper and cite their sources.
May warrant caution with large pectoral ICDs and certain leads; leaves open prospective comparison of modern generator sizes and systems.
As the majority of ICDs with transvenous leads are now implanted in the pectoral region, complications associated with the technique are being identified. To determine the incidence of lead complications in patients with transvenous defibrillator leads and ICDs implanted in the pectoral region, 132 unselected consecutive patients with transvenous defibrillator leads had ICDs implanted in the pectoral region. Three lead systems were used: (1) lead system 1 (45 patients) consisted of a transvenous pacing sensing lead and a superior vena cava coil with a submuscular patch used for defibrillation; (2) lead system 2 (36 patients) utilized a CPI Endotak lead system; and (3) lead system 3 (51 patients) utilized a Medtronic Transvene lead system. Patients were followed for 3-54 months (cumulative 2,269, mean 18 months). The average duration of follow-up with the three systems was 32, 12 and 11 months, respectively. At 30 months follow-up, all three lead systems had a low incidence of complications. However, there was a 13% overall incidence (45% actuarial incidence) of erosion of the insulation of the pacing sensing lead of system 1 at 50 months of follow-up. All lead complications were seen in patients with ICDs whose weights were > 195 g and volumes > 115 cc. The erosion was probably a consequence of the pressure by the large ICD against the lead in the pectoral pocket. Follow-up with lead systems 2 and 3 is relatively short (average 12 months) but no lead erosions were seen. Pectoral implantation of ICDs with long transvenous leads and large generators is associated with a moderate risk of late complications in the form of insulation breaks caused by pressure of the generator against the leads. The use of less redundant leads coupled with smaller ICDs will probably eliminate this complication.
No takes yet. Share an insight, caveat, or question.
Mehta et al. (1998) conducted a cohort in Pectoral defibrillator implants with transvenous leads (n=132). Pectoral ICD implantation with transvenous leads was evaluated on Lead complications (insulation erosion). Pectoral implantation of large ICDs with transvenous leads resulted in a 13% overall incidence of insulation erosion in one lead system at 50 months of follow-up.