Key result
Deep negative unipolar His-bundle EGM deflection linked to ~50% lower HBP threshold versus positive EGM.
Why the study?
High capture thresholds are common in His-bundle pacing, but few electrophysiological predictors of low capture threshold have been determined.
Does a deep negative deflection in unipolar His-bundle electrogram predict lower pacing thresholds in patients undergoing His-bundle pacing for bradycardia?
Observational (n=51)
Does a deep negative deflection in unipolar His-bundle electrogram predict lower pacing thresholds in patients undergoing His-bundle pacing for bradycardia?
A deep negative deflection (≥0.060 mV) in unipolar His-bundle electrograms at fixation predicts excellent short-term and permanent His-bundle pacing thresholds.
Answers are grounded in this paper and cite their sources.
May inform His-bundle lead positioning for lower thresholds; hypothesis-generating and requires prospective validation before practice change.
Background His-bundle pacing (HBP) is a physiological form of pacing. Although high capture thresholds are common, few predictors of low HBP threshold have been determined. We aimed to identify electrophysiological predictors. Methods Fifty-one patients (53% with atrioventricular block) underwent HBP for bradycardia with an intrinsic QRS duration of <120 ms. Attempts to anchor the HBP lead were guided by unipolar His-bundle electrograms (HB EGMs) recorded with an electrophysiology recording system. Patients were followed-up for >6 months. Results In total, 153 attempts at anchoring the HBP lead were made, of which, 45 achieved acceptable HBP thresholds (≤2.5 V at 1 ms). The amplitude of negative deflection in HB EGM and the selective HBP form at fixation were independently associated with achieving an acceptable threshold. A negative amplitude of ≥0.060 mV in HB EGM was determined as the optimal value for identifying the acceptable threshold. This deep negative HB EGM was recorded with an HBP threshold of 1.4±1.3 V (in 34 attempts), significantly lower than that of positive HB EGM without deep negative deflection (2.8±1.3 V, in 31 trials; or >5 V, in 38 trials). The permanent HBP lead remained with deep negative (≥0.060 mV) or positive HB EGMs in 28 and 14 patients, respectively, and with positive or negative HB injury current in 19 and 23 patients, respectively. During follow-up, increased HBP threshold of >1 V was significantly more prevalent in the positive HB EGM group. The HBP thresholds of deep negative HB EGM and HB injury current, but not of the selective HBP group, were significantly lower than the other subgroups during follow-up. Conclusions Deep negative HB EGM at fixation was associated with an excellent short-term HBP threshold, similar to HB injury current. Analysis of unipolar HB EGM postfixation may enable prediction of permanent HBP threshold.
No takes yet. Share an insight, caveat, or question.
Sato et al. (2019) conducted an observational in Bradycardia requiring His-bundle pacing (n=51). Deep negative deflection (≥0.060 mV) in unipolar His-bundle electrogram vs. Positive His-bundle electrogram without deep negative deflection was evaluated on His-bundle pacing threshold. Deep negative deflection in unipolar His-bundle electrogram at fixation was associated with a lower HBP threshold compared to positive HB EGM.
Synapse has enriched 10 related papers. Consider them for comparative context: