Key result
Internal atrial defibrillation is feasible in humans using <2 J biphasic shocks with right atrium and coronary sinus electrodes, without causing significant pathological damage at <5 J.
Why the study?
Are implantable atrial defibrillators a feasible and safe therapy for patients with symptomatic, paroxysmal atrial fibrillation?
Are implantable atrial defibrillators a feasible and safe therapy for patients with symptomatic, paroxysmal atrial fibrillation?
Implantable atrial defibrillators represent a feasible nonpharmacologic option for paroxysmal atrial fibrillation, though further safety and tolerability improvements are needed.
Answers are grounded in this paper and cite their sources.
Supports feasibility of implantable atrial defibrillators for paroxysmal AF; leaves open long-term safety, tolerability, and need for prospective trials.
Due to the limited efficacy of antiarrhythmic drugs for atrial fibrillation, several nonpharmacologic therapeutic options have evolved. One of these is an implantable atrial defibrillator. Recent studies have shown that internal atrial defibrillation is feasible with relatively low energies. To date, the optimal electrode configuration involves large surface area catheters in the right atrium and coronary sinus. In humans, atrial defibrillation can generally be achieved with < 2 J using this electrode configuration and a biphasic shock waveform. For shocks < 5 J, there is no significant pathological damage to the atria or coronary sinus. Further investigation is needed to guarantee that atrial defibrillation shocks do not provoke ventricular arrhythmias. Preliminary data suggest that atrial defibrillation shocks synchronized to R waves that are not closely coupled are safe. In addition, the shocks are well tolerated if the shock energy is < 1.5 J. With additional studies to confirm the safety of implantable atrial defibrillators, further reduce shock energy, and improve patient tolerance, an implantable atrial defibrillator can become an acceptable therapy for patients with symptomatic, paroxysmal atrial fibrillation.
No takes yet. Share an insight, caveat, or question.
Hillsley et al. (1995) conducted a review in Atrial fibrillation. Implantable atrial defibrillator was evaluated. Internal atrial defibrillation is feasible in humans using <2 J biphasic shocks with right atrium and coronary sinus electrodes, without causing significant pathological damage at <5 J.