Fast-SENC cardiac magnetic resonance imaging provided highly reproducible measurements of left ventricular global longitudinal strain (ICC 0.94) and global circumferential strain (ICC 0.95) between studies.
Observational (n=18)
Blinded observers
No
Heart failure and healthy volunteers (n=18)
fast-SENC cardiac magnetic resonance imaging
Interstudy reproducibility (Intraclass correlation coefficient) for Global Longitudinal Strain (0.84 to 0.98)
Myocardial strain is a well validated parameter for estimating left ventricular (LV) performance. The aim of our study was to evaluate the inter-study as well as intra- and interobserver reproducibility of fast-SENC derived myocardial strain. Eighteen subjects (11 healthy individuals and 7 patients with heart failure) underwent a cardiac MRI examination including fast-SENC acquisition for evaluating left ventricular global longitudinal (GLS) and circumferential strain (GCS) as well as left ventricular ejection fraction (LVEF). The examination was repeated after 63 range 49‒87 days and analyzed by two experienced observers. Ten datasets were repeatedly assessed after 1 month by the same observer to test intraobserver variability. The reproducibility was measured using the intraclass correlation coefficient (ICC) and Bland-Altman analysis. Patients with heart failure demonstrated reduced GLS and GCS compared to healthy controls (-15.7 ± 3.7 vs. -20.1 ± 1.4; p = 0.002 for GLS and -15.3 ± 3.7 vs. -21.4 ± 1.1; p = 0.001 for GCS). The test-retest analysis showed excellent ICC for LVEF (0.92), GLS (0.94) and GCS (0.95). GLS exhibited excellent ICC (0.99) in both intra- and interobserver variability analysis with very narrow limits of agreement (-0.6 to 0.5 for intraobserver and -1.3 to 0.96 for interobserver agreement). Similarly, GCS showed excellent ICC (0.99) in both variability analyses with narrow limits of agreement (-1.1 to 1.2 for intraobserver and -1.7 to 1.3 for interobserver agreement), whereas LVEF showed larger limits of agreement (-14.4 to 10.1). The analysis of fast-SENC derived myocardial strain using cardiac MRI provides a highly reproducible method for assessing LV functional performance.
Building similarity graph...
Analyzing shared references across papers
Loading...
Giuşcă et al. (Fri,) conducted a observational in Heart failure and healthy volunteers (n=18). fast-SENC cardiac magnetic resonance imaging was evaluated on Interstudy reproducibility (Intraclass correlation coefficient) for Global Longitudinal Strain (95% CI 0.84 to 0.98). Fast-SENC cardiac magnetic resonance imaging provided highly reproducible measurements of left ventricular global longitudinal strain (ICC 0.94) and global circumferential strain (ICC 0.95) between studies.
synapsesocial.com/papers/6a0f6ec89e54838161fcb6af — DOI: https://doi.org/10.1038/s41598-018-32226-3
Sorin Giuşcă
Gerresheimer (Germany)
Grigorios Korosoglou
Cardiac Imaging
Victoria Zieschang
Deutsches Herzzentrum der Charité
Scientific Reports
University of Chicago
The University of Sydney
Charité - Universitätsmedizin Berlin
Building similarity graph...
Analyzing shared references across papers
Loading...