Key result
Free-breathing ferumoxytol-enhanced UTE-MRA matches conventional gadolinium-enhanced MRA for pulmonary vasculature image quality.
Why the study?
To evaluate the feasibility of ferumoxytol-enhanced UTE-MRA for depiction of pulmonary vascular and nonvascular structures.
Does ferumoxytol-enhanced UTE-MRA provide comparable or superior image quality of pulmonary vascular and nonvascular structures compared to gadolinium-enhanced conventional MRA in healthy volunteers?
Observational (n=20)
Does ferumoxytol-enhanced UTE-MRA provide comparable or superior image quality of pulmonary vascular and nonvascular structures compared to gadolinium-enhanced conventional MRA in healthy volunteers?
Free-breathing ferumoxytol-enhanced UTE-MRA is a feasible alternative to gadolinium-enhanced conventional MRA for simultaneous assessment of pulmonary vasculature and nonvascular structures.
Answers are grounded in this paper and cite their sources.
Comparable quality supports ferumoxytol UTE-MRA feasibility in volunteers; leaves open validation in patients.
Purpose To evaluate the feasibility of ferumoxytol (FE)‐enhanced UTE‐MRA for depiction of the pulmonary vascular and nonvascular structures. Methods Twenty healthy volunteers underwent contrast‐enhanced pulmonary MRA at 3 T during 2 visits, separated by at least 4 weeks. Visit 1: The MRA started with a conventional multiphase 3D T 1 ‐weighted breath‐held spoiled gradient‐echo MRA before and after the injection of 0.1 mmol/kg gadobenate dimeglumine (GD). Subsequently, free‐breathing GD‐UTE‐MRA was acquired as a series of 3 flip angles (FAs) (6°, 12°, 18°) to optimize T 1 weighting. Visit 2: After the injection of 4 mg/kg FE, MRA was performed during the steady state, starting with a conventional 3D T 1 ‐weighted breath‐held spoiled gradient‐echo MRA and followed by free‐breathing FE‐UTE‐MRA, both at 4 different FAs (6°, 12°, 18°, 24°). The optimal FA for best T 1 contrast was evaluated. Image quality at the optimal FA was compared between methods on a 4‐point ordinal scale, using multiphase GD conventional pulmonary MRA (cMRA) as standard of reference. Results Flip angle in the range of 18°‐24° resulted in best T 1 contrast for FE cMRA and both UTE‐MRA techniques ( p > .05). At optimized FA, image quality of the vasculature was good/excellent with both FE‐UTE‐MRA and GD cMRA (98% versus 97%; p = .51). Both UTE techniques provided superior depiction of nonvascular structures compared with either GD‐enhanced or FE‐enhanced cMRA ( p < .001). However, GD‐UTE‐MRA showed the lowest image quality of the angiogram due to low image contrast. Conclusion Free‐breathing UTE‐MRA using FE is feasible for simultaneous assessment of the pulmonary vasculature and nonvascular structures. Patient studies should investigate the clinical utility of free‐breathing UTE‐MRA for assessment of pulmonary emboli.
No takes yet. Share an insight, caveat, or question.
Knobloch et al. (2019) conducted an observational in Healthy volunteers (n=20). Ferumoxytol-enhanced UTE-MRA vs. Gadobenate dimeglumine-enhanced conventional MRA was evaluated on Image quality of the vasculature (good/excellent) (p=0.51). Free-breathing ferumoxytol-enhanced UTE-MRA provided good/excellent image quality of the pulmonary vasculature comparable to gadolinium-enhanced conventional MRA.
Synapse has enriched 10 related papers. Consider them for comparative context: