Why the study?
The influence of cuff inflations on nighttime measurements during 24 h ambulatory blood pressure measurements was unknown.
Does cuff inflation during 24-hour ambulatory blood pressure measurement affect sleep parameters and blood pressure compared to cuffless measurement?
Does cuff inflation during 24-hour ambulatory blood pressure measurement affect sleep parameters and blood pressure compared to cuffless measurement?
Cuff inflations during 24-hour ambulatory blood pressure monitoring do not significantly impact sleep blood pressure, sleep efficiency, or desaturations compared to cuffless monitoring.
Answers are grounded in this paper and cite their sources.
Cuff-based ABPM was not associated with worse asleep BP or efficiency than cuffless; leaves open whether arousal differences warrant device changes.
The influence of cuff inflations on night-time measurements during 24 h ambulatory blood pressure (BP) measurements is unknown. We investigated the potential effect of cuff inflations on sleep parameters using measurements taken simultaneously with a cuffless device using pulse-transit-time (PTT). On the first day of measurement, standard cuff-based 24 h BP and cuffless measurements were simultaneously performed on the right and left arms (CUFF/PTT-D). In this experiment, 1–2 days after the first measurement, the cuffless device was worn alone (PTT-D). Only data from the cuffless device were analyzed. The following mean sleep parameters were analyzed: mean systolic and diastolic BP, arousals, sleep efficiency, total arousals, arousal per hour, and desaturations. In total, 21 individuals were prospectively enrolled. The mean (SD) age was 47 (±15) years, and 57% were female. The mean systolic asleep BP during CUFF/PTT-D and during PTT-D were 131 (±21) and 131 (±26) mmHg, respectively. The mean diastolic asleep BP values during CUFF/PTT-D and during PTT-D were 80 (±14) and 84 (±14) mmHg, respectively (p = 0.860, p = 0.100, respectively). Systolic and diastolic asleep mean difference was 0.1 (±18.0) and −3.6 (±9.8) mmHg, respectively. There were significantly more total arousals during PTT-D (p = 0.042). There were no significant differences seen in sleep efficiency (p = 0.339) or desaturations (p = 0.896) between the two measurement periods. We could not show any significant impact from cuff inflations during sleep, as documented by PTT-D measurements.
No takes yet. Share an insight, caveat, or question.
Socrates et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: