Key result
Doppler echocardiography was inaccurate (defined as >10 mm Hg difference from invasive measurement) in 48% of cases when estimating pulmonary artery systolic pressure compared to right-heart catheterization.
Why the study?
Does Doppler echocardiography accurately estimate pulmonary artery pressure and cardiac output compared to right-heart catheterization in patients with pulmonary hypertension?
Observational (n=65)
Single-blind
No
Does Doppler echocardiography accurately estimate pulmonary artery pressure and cardiac output compared to right-heart catheterization in patients with pulmonary hypertension?
Doppler echocardiography frequently provides inaccurate estimates of pulmonary artery pressure and cardiac output compared to invasive right-heart catheterization in patients with pulmonary hypertension.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Fisher et al. (2009) conducted an observational in Pulmonary Hypertension (n=65). Doppler echocardiography vs. Right-heart catheterization was evaluated on Accuracy of Doppler echocardiography for estimating pulmonary artery systolic pressure. Doppler echocardiography was inaccurate (defined as >10 mm Hg difference from invasive measurement) in 48% of cases when estimating pulmonary artery systolic pressure compared to right-heart catheterization.