Why the study?
Lesion morphologies from longer radiofrequency durations with the novel DiamondTemp ablation system had not been previously described.
Does varying application durations and temperature set-points with the DiamondTemp ablation system affect lesion morphology in an ex vivo porcine model?
Population
280 applications in a bench model using porcine myocardium
Comparison
Temperature set-points of 50 °C vs 60 °C across randomized durations of 5–120 s
Design
Randomized bench model study
Authors
Loading...
Ex vivo lesion dimensions plateau after 60 s; leaves open optimal DiamondTemp dosing in humans.
Does varying application durations and temperature set-points with the DiamondTemp ablation system affect lesion morphology in an ex vivo porcine model?
The DiamondTemp ablation system rapidly creates wide lesions with depth plateauing after 60 seconds, suggesting that a 'Time after Temperature 60 °C' approach may optimize ablation dosing.
Dhanjal et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: