Why the study?
Accurate distal contact force measurement was needed by developing a temperature-compensated sensor integrated into an AF ablation catheter.
A novel FBG-based temperature-compensated contact force sensor for AF ablation catheters demonstrates high sensitivity and resolution in bench testing, suggesting suitability for industrial mass production.
Enables AF ablation catheter prototype development; leaves open human validation and clinical outcomes.
OBJECTIVE: To accurately achieve distal contact force, a novel temperature-compensated sensor is developed and integrated into an atrial fibrillation (AF) ablation catheter. METHODS: A dual elastomer-based dual FBGs structure is used to differentiate the strain on the two FBGs to achieve temperature compensation, and the design is optimized and validated by finite element simulation. RESULTS: The designed sensor has a sensitivity of 90.5 pm/N, resolution of 0.01 N, and root-mean-square error (RMSE) of 0.02 N and 0.04 N for dynamic force loading and temperature compensation, respectively, and can stably measure distal contact forces with temperature disturbances. CONCLUSION: Due to the advantages, i.e., simple structure, easy assembly, low cost, and good robustness, the proposed sensor is suitable for industrial mass production.
No takes yet. Share an insight, caveat, or question.
Lou et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: