Key result
Fiber Bragg grating sensors map myocardial radiofrequency ablation temperatures, capturing gradients up to ~34 °C.
Why the study?
Current RFA catheters provide only single-point temperature measurements within the probe, lacking information regarding myocardial tissue temperature changes to help predict lesion dimensions and prevent tissue damage.
Fiber Bragg grating sensors demonstrate feasibility for real-time, high-resolution 3D temperature mapping within myocardial tissue during radiofrequency ablation in an ex vivo model.
Radiofrequency ablation (RFA) is the most widely used technique for the treatment of cardiac arrhythmias. A variety of factors, such as the electrode tip shape, the force exerted on the tissue by the catheter and the delivered power, combine to determine the temperature distribution, and as consequence, the lesion shape and size. In this context, being able to know the temperature reached in the myocardium during the RFA can be helpful for predicting the lesion dimensions to prevent the occurrence of undesired tissue damage. The catheters used so far in such procedures provide single-point temperature measurements within the probe (by means of embedded thermocouples or thermistors), so no information regarding the temperature changes occurring in myocardial tissues can be retrieved. The aim of this study was to assess the feasibility of fiber Bragg grating sensors (FBGs) to perform multi-point and millimetric-scale temperature measurements within myocardium subjected to RFA. The assessment has been performed on ex vivo porcine myocardium specimens undergoing RFA. Data show the feasibility of the proposed solution in providing spatial temperature distribution within the myocardial tissue during the entire RFA. These high-resolved measurements may allow reconstructing the temperature distribution in the tissue. This study lays the foundations for the implementation of 3D thermal maps to investigate how the supplied power, treatment time, force of contact and irrigation flow of the catheter influence the thermal effects within the tissue.
No takes yet. Share an insight, caveat, or question.
Zaltieri et al. (2020) studied Cardiac arrhythmias (RFA treatment) (n=2). Fiber Bragg Grating (FBG) sensors was evaluated on Feasibility of multi-point temperature measurements. Fiber Bragg grating sensors successfully provided high-resolution spatial temperature measurements within ex vivo porcine myocardium during radiofrequency ablation, recording variations up to 33.5 °C.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: