Why the study?
Steam pops are a rare complication of radiofrequency ablation that are difficult to predict, prompting an assessment of how electrode-tissue coverage influences steam pop incidence and lesion size.
Does the level of coverage between the RF ablation electrode and cardiac tissue influence steam pop incidence and lesion size in an ex vivo porcine model?
Does the level of coverage between the RF ablation electrode and cardiac tissue influence steam pop incidence and lesion size in an ex vivo porcine model?
Increased electrode-tissue coverage during RF ablation significantly increases both lesion size and the risk of steam pops.
High electrode-tissue coverage may raise steam pop risk in RF ablation; leaves open optimal parameters for clinical translation.
Background Steam pops are a rare complication associated with radiofrequency (RF) ablation and are hard to predict. The aim of this study was to assess the influence of coverage between the RF ablation electrode and cardiac tissue on steam pop incidence and lesion size. Methods and results An ex vivo model using porcine cardiac preparations and contact force sensing catheters was designed to perform RF ablations at different coverage levels between the RF electrode and cardiac tissue. During coverage level I, only the distal part of the ablation electrode was in contact with tissue. During coverage level II half of the ablation electrode, and during coverage level III the entire ablation electrode was embedded in tissue. RF applications ( n = 60) at different coverage levels I–III were systematically performed using the same standardized ablation protocol. Ablations during coverage level III resulted in a significantly higher rate of steam pops (100%) when compared to ablations during coverage level II (10%) and coverage level I (0%), log rank p < .001. Coverage level I ablations resulted in significantly smaller lesion depths, diameters, and impedance drops when compared to higher coverage level ablations, p < .001. In the controlled ex vivo model, there was no difference in applied contact force or energy between different coverage levels. Conclusions The level of coverage between RF electrode, cardiac tissue, and the surrounding fluid significantly influenced the incidence of steam pops in an ex vivo setup. Larger coverage between RF electrode and tissue resulted in significantly larger lesion dimensions.
No takes yet. Share an insight, caveat, or question.
Bourier et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: