Key result
Radiofrequency catheter ablation using closed loop temperature control resulted in a lower incidence of coagulum formation compared to power control mode (0.8% vs 2.2%, P<0.01).
Why the study?
Does temperature-controlled radiofrequency ablation reduce the incidence of coagulum compared to power-controlled ablation in patients undergoing catheter ablation?
Population
270 patients undergoing catheter ablation of atrioventricular nodal reentrant tachycardia, an accessory…
Comparison
Radiofrequency catheter ablation in temperature… vs Radiofrequency catheter ablation in power…
Design
Cohort
Authors
Loading...
Temperature control was associated with less coagulum; hypothesis-generating and requires randomized confirmation before practice change.
Cohort (n=270)
Yes
Does temperature-controlled radiofrequency ablation reduce the incidence of coagulum compared to power-controlled ablation in patients undergoing catheter ablation?
Absolute Event Rate: 0.8% vs 2.2%
p-value: p=< .01
Closed loop temperature control during radiofrequency ablation significantly reduces coagulum formation compared to fixed power control.
Calkins et al. (1994) conducted a cohort in Atrioventricular nodal reentrant tachycardia, accessory pathway, and/or atrioventricular junction (n=270). Radiofrequency catheter ablation in temperature control mode vs. Radiofrequency catheter ablation in power control mode was evaluated on Incidence of developing a coagulum during radiofrequency applications (p=< .01). Radiofrequency catheter ablation using closed loop temperature control resulted in a lower incidence of coagulum formation compared to power control mode (0.8% vs 2.2%, P<0.01).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: