Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium.

View Full Paper
DHDavid E. HainesAVAnthony F. Verow

Key Result

Radiofrequency ablation with peak temperatures >100°C caused a sudden rise in electrical impedance in 80% of in vivo lesions compared to 6.25% with temperatures <100°C (p=0.0001).

Structured PICO

Does maintaining electrode-tissue interface temperature <100 degrees C prevent sudden rise in electrical impedance during radiofrequency ablation in a canine model?

P
Population
Canine model, in vitro (n=49) and in vivo (n=31) lesions
I
Intervention
Radiofrequency ablation with varying power settings
C
Comparator
Ablations with peak electrode-tissue interface temperature <100 degrees C versus >100 degrees C
O
Outcome
Changes in electrical impedance correlated with electrode-tissue interface temperaturesurrogate

Maintaining electrode-tissue interface temperature below 100°C during radiofrequency ablation prevents sudden rises in electrical impedance and associated complications like thrombus formation.

Main Result

Absolute Event Rate: 80% vs 6.25%

p-value: p=0.0001

Abstract

The purpose of this study was to correlate changes in electrical impedance with the electrode-tissue interface temperature and to characterize the associated events occurring at the catheter tip electrode. In a canine model, lesions were created in vitro (n = 49) and in vivo (n = 31) and radiofrequency power settings were varied. Electrode-tissue interface temperature, delivered current, and voltage were recorded, and impedance was calculated. A sudden rise in electrical impedance was seen in only two of 17 ablations in vitro and in one of 16 ablations in vivo with a peak electrode-tissue interface temperature of less than 100 degrees C compared with 29 of 32 ablations in vitro (p = 0.0001) and 12 of 15 ablations in vivo with a temperature of more than 100 degrees C (p = 0.0001). This phenomenon was associated with the observation of boiling and popping at the tip in in vitro preparations and tissue avulsion and thrombus formation on the catheter tip in in vivo studies. The lesion size was directly proportional to the peak temperature for all ablations but not to the peak power, current, or voltage during radiofrequency catheter ablation in the heart. Maintaining electrode-tissue interface temperature at less than 100 degrees C during radiofrequency catheter ablation in the heart may avoid the complications associated with the sudden rise in electrical impedance.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Haines et al. (1990) studied Ventricular myocardium undergoing radiofrequency ablation (n=80). Radiofrequency ablation with peak temperature >100 degrees C vs. Radiofrequency ablation with peak temperature <100 degrees C was evaluated on Sudden rise in electrical impedance (p=0.0001). Radiofrequency ablation with peak temperatures >100°C caused a sudden rise in electrical impedance in 80% of in vivo lesions compared to 6.25% with temperatures <100°C (p=0.0001).

synapsesocial.com/papers/6a0ff561fb2817e31dfcd825https://doi.org/10.1161/01.cir.82.3.1034