Key result
Pre-ablation measurement of impedance and temperature rise (DeltaIMP and DeltaT) significantly correlated with the obtained lesion size during radiofrequency ablation in vitro (P<0.001).
Why the study?
Can pre-ablation measurements of impedance and temperature rise indicate the resulting lesion size during radiofrequency ablation?
Population
In vitro strips of left ventricular porcine myocardium
Design
Preclinical
Authors
Loading...
Should not guide clinical ablation decisions yet; leaves open translation of in vitro lesion prediction to in vivo use.
Can pre-ablation measurements of impedance and temperature rise indicate the resulting lesion size during radiofrequency ablation?
p-value: p=<0.001
Pre-ablation measurements of impedance and temperature rise can predict lesion size during radiofrequency ablation in an in vitro model.
STAGEGAARD et al. (2005) studied Arrhythmias. Pre-ablation measurement of DeltaIMP and DeltaT was evaluated on Correlation between the product of DeltaT and DeltaIMP and obtained lesion size (p=<0.001). Pre-ablation measurement of impedance and temperature rise (DeltaIMP and DeltaT) significantly correlated with the obtained lesion size during radiofrequency ablation in vitro (P<0.001).