Key result
Radiofrequency ablation using large-tip catheters produced lesion sizes that varied with blood flow rate, unlike cooled-tip catheters which were unaffected.
Why the study?
Does convective cooling flow rate differentially affect lesion size created by cooled-tip versus large-tip radiofrequency ablation catheters in porcine myocardium?
RCT (n=72)
Randomized
Does convective cooling flow rate differentially affect lesion size created by cooled-tip versus large-tip radiofrequency ablation catheters in porcine myocardium?
Blood flow rate significantly affects lesion size for large-tip but not cooled-tip RF ablation catheters, with cooled-tip creating larger lesions at low flow and large-tip creating larger lesions at high flow.
Answers are grounded in this paper and cite their sources.
Caution with large-tip catheters in variable-flow regions; extends preclinical data but leaves optimal selection open pending clinical studies.
BACKGROUND: Both actively cooled-tip and large-tip catheters are currently available clinically to create large endomyocardial lesions during application of radiofrequency (RF) energy. The purpose of this study was to compare the effect of convective cooling at physiologic flow rates on RF lesion size using both actively cooled and large-tip catheters. METHODS: Porcine hearts were sectioned into 72 pieces and placed in a temperature-controlled saline bath (37 degrees C) with varying directed flow rates (0, 1, 2, and 3 L/min). Cooled-tip RF ablation (4 mm tip) was performed for 1 minute on 36 tissue sections with power manually titrated to keep tip temperature below 40 degrees C. Large-tip catheter ablation (10 mm tip) was performed at 65 degrees C target temperature for 1 minute on 36 tissue sections. For each catheter, flow rates were randomized between applications. The tissue pieces were sectioned and measured to determine lesion depth, width, and volume. RESULTS: Lesion dimensions were independent of the flow rate for the cooled-tip catheter (mean volumes: 382.0 +/- 121.6, 419.9 +/- 133.4, 375.9 +/- 169.1, and 346.7 +/- 173.4 mm(3) for 0, 1, 2, and 3 L/min flow rate, respectively, P = 0.78). For the large-tip catheter, lesion size varied significantly with flow, such that higher flow rates produced larger lesions (mean volumes: 120.7 +/- 50.7, 256.5 +/- 97.9, 393.4 +/- 149.9, and 548.9 +/- 157.0 mm(3) for 0, 1, 2, and 3 L/min flow rate respectively, P < 0.001) CONCLUSION: During RF ablation, blood flow rate significantly affects lesion size for large-tip but not cooled-tip catheters. At low flow rates (0-1 L/min) cooled-tip catheters create larger lesions, while at high flow rates (3 L/min) large-tip catheters create larger lesions.
No takes yet. Share an insight, caveat, or question.
Pilcher et al. (2006) reported an RCT. Cooled-tip catheter radiofrequency ablation vs. Large-tip catheter radiofrequency ablation was evaluated on Lesion volume (p=<0.001). Radiofrequency ablation using large-tip catheters produced lesion sizes that varied with blood flow rate, unlike cooled-tip catheters which were unaffected.