Key result
ADAPT-HEART algorithm detects structural heart disease from single-lead ECGs, predicting ~4-fold higher incident risk.
Why the study?
AI-enhanced 12-lead ECG can detect structural heart diseases but has a limited role in community-based screening.
Does an ensemble deep learning algorithm using single-lead ECGs accurately detect and predict structural heart disease in a diverse population?
Observational (n=99,205)
Yes
Does an ensemble deep learning algorithm using single-lead ECGs accurately detect and predict structural heart disease in a diverse population?
Effect estimate: AUROC 0.879 (95% CI 0.870-0.888)
A noise-resilient deep learning algorithm applied to single-lead ECGs can accurately detect and predict structural heart disease, providing a scalable strategy for community-based screening.
May support scalable single-lead ECG screening; hypothesis-generating pending prospective trials.
Background and Aims: AI-enhanced 12-lead ECG can detect a range of structural heart diseases (SHDs) but has a limited role in community-based screening. We developed and externally validated a noise-resilient single-lead AI-ECG algorithm that can detect SHD and predict the risk of their development using wearable/portable devices. Methods: Using 266,740 ECGs from 99,205 patients with paired echocardiographic data at Yale New Haven Hospital, we developed ADAPT-HEART, a noise-resilient, deep-learning algorithm, to detect SHD using lead I ECG. SHD was defined as a composite of LVEF<40%, moderate or severe left-sided valvular disease, and severe LVH. ADAPT-HEART was validated in four community hospitals in the US, and the population-based cohort of ELSA-Brasil. We assessed the model's performance as a predictive biomarker among those without baseline SHD across hospital-based sites and the UK Biobank. Results: The development population had a median age of 66 [IQR, 54-77] years and included 49,947 (50.3%) women, with 18,896 (19.0%) having any SHD. ADAPT-HEART had an AUROC of 0.879 (95% CI, 0.870-0.888) with good calibration for detecting SHD in the test set, and consistent performance in hospital-based external sites (AUROC: 0.852-0.891) and ELSA-Brasil (AUROC: 0.859). Among those without baseline SHD, high vs. low ADAPT-HEART probability conferred a 2.8- to 5.7-fold increase in the risk of future SHD across data sources (all P<0.05). Conclusions: We propose a novel model that detects and predicts a range of SHDs from noisy single-lead ECGs obtainable on portable/wearable devices, providing a scalable strategy for community-based screening and risk stratification for SHD.
No takes yet. Share an insight, caveat, or question.
Aminorroaya et al. (2024) conducted an observational in Structural Heart Disease (n=99,205). ADAPT-HEART (AI-ECG algorithm) was evaluated on Discrimination for detecting the presence of composite structural heart disease (AUROC) (AUROC 0.879, 95% CI 0.870-0.888). The ADAPT-HEART deep learning algorithm detected structural heart disease from single-lead ECGs with an AUROC of 0.879 and predicted a 4-fold higher hazard of incident disease.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: