This is the authors' abstract. We don't add key points for this paper.
We present an experimental method for performing direct, high-speed dynamic pressure measurements of MEMS-based sensors with bandwidths exceeding 1 MHz. A focused nanosecond-pulsed laser beam is used to generate a laser-induced spark, which rapidly expands into a supersonic shock wave that excites the pressure transducer. A windowed chamber with an integrated heating element enables characterization of the sensor's dynamic pressure response at temperatures ranging from 25°C to 1000°C. A precision PID-controlled pressure regulation system is employed to establish static pressure setpoints from 1 Pa to 33,000 Pa (absolute), which are then dynamically modulated by the laser-induced shock wave. This method offers significant advantages in repeatability, accuracy, and bandwidth compared to other direct and indirect measurement techniques, such as shock tubes, ultrasonic excitation, or low-frequency reciprocal calibration.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Lambert et al. (2025) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: