Key points are not available for this paper at this time.
We have studied the influence of triglyceride-rich particles on the analytical bias of apolipoprotein B measurements by various immunoturbidimetric methods. Three commercially available methods grossly overestimate apolipoprotein B in samples with even moderately above-normal triglyceride concentrations. This effect is due to the increased relative reactivity of very-low-density lipoproteins in these reagent systems, and can be eliminated by including Tween 20 (2 g/L) in the reagent buffer. We have developed, and describe, an automated immunoturbidimetric method that allows the accurate determination of apolipoprotein B in the plasma of patients with hypertriglyceridemia.
Building similarity graph...
Analyzing shared references across papers
Loading...
T A Walmsley
Canterbury Health Laboratories
Susan Grant
Princess Margaret Hospital
Peter M. George
Christchurch Hospital
Clinical Chemistry
Princess Margaret Hospital
Building similarity graph...
Analyzing shared references across papers
Loading...
Walmsley et al. (Wed,) studied this question.
synapsesocial.com/papers/6a10436b5725bbd5cc60bf3f — DOI: https://doi.org/10.1093/clinchem/37.5.748