This is the authors' abstract. We don't add key points for this paper.
We have designed a re-usable dialysis cell and a complex dialysis buffer, with which undiluted serum samples can be dialyzed with minimal changes in their serum matrix. Dialysate thyroxin (free T4) is then measured by a sensitive RIA for T4. The range of reportability was 2-128 ng/L, the normal range was 8-27 ng/L, and the interassay CV was 7%. Free T4 concentrations in various disorders were as follows: hyperthyroidism, 32-478 ng/L; in both excess thyroxin-binding globulin (TBG) and familial dysalbuminemic hyperthyroxinemia, 9-27 ng/L; primary hypothyroidism, less than 2-7 ng/L; central hypothyroidism, 4-6 ng/L; severe TBG deficiency, 9-25 ng/L; hypothyroxinemias of nonthyroidal illness, 8-35 ng/L. With this free-T4 assay, which is adaptable to clinical laboratory use, one can differentiate hyperthyroidism from the major euthyroid hyperthyroxinemias and hypothyroidism from the major euthyroid hypothyroxinemias.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Nelson et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: