Key result
Room temperature FT4 assays yield ~2-fold higher values than reference methods in low thyroxine-binding globulin samples.
Why the study?
Does performing free thyroxine (FT4) assays at room temperature affect measurement accuracy compared to 37 degrees C?
Does performing free thyroxine (FT4) assays at room temperature affect measurement accuracy compared to 37 degrees C?
Performing free thyroxine assays at room temperature is fundamentally incorrect and causes falsely elevated values in samples with low thyroxine-binding globulin.
Room-temperature two-step FT4 assays warrant caution in low-TBG samples; leaves open need for 37°C standardization across labs.
Two recently developed two-step, or "back-titration" assay kits for free thyroxine (FT4)--one based on a europium-labeled derivative of T4, the other on conventional radiolabeled T4--were compared with both symmetrical dialysis and an indirect FT4 radioimmunoassay. Discrepancies between the two-step and the other two methods were observed, particularly in samples that had very low to zero thyroxine-binding globulin (TBG) content. In those instances the two-step methods gave values almost twice as high as the other methods. This effect could be largely reversed in one of the two-step assays and completely reversed in the other by performing the first incubation step at 37 degrees C rather than at room temperature, as prescribed by the manufacturers. When symmetrical dialysis is performed at both temperatures, FT4 at room temperature is about 40% of the amount determined at 37 degrees C, except in zero-TBG samples, where it averages almost 80% of the value at 37 degrees C. Moreover, we demonstrated that the affinity of TBG for T4 is much more temperature-dependent than the affinity of transthyretin and albumin for T4, so that the net temperature effect on the FT4 in a sample depends on the relative contribution of TBG to total binding. We conclude that performing FT4 assays at room temperature is principally incorrect and leads to falsely increased values when samples with very low TBG concentrations are analyzed.
No takes yet. Share an insight, caveat, or question.
Ross et al. (1992) studied Free thyroxine (FT4) measurement. Two-step FT4 assays at room temperature vs. Symmetrical dialysis and indirect FT4 radioimmunoassay at 37 degrees C was evaluated on FT4 measurement accuracy. Performing two-step FT4 assays at room temperature yielded values almost twice as high as reference methods in samples with very low thyroxine-binding globulin, an effect reversed at 37 degrees C.