Key result
Veterinary ECGs measure R-wave amplitudes ~16% higher than human ECGs, identifying more canine ventricular enlargement.
Why the study?
To compare measured ECG intervals and amplitudes obtained using electrocardiographs adapted for veterinary versus human use and to establish ECG reference intervals for endurance-trained dogs.
Does a veterinary digital electrocardiograph produce different ECG values and reference intervals compared to a human digital electrocardiograph in endurance-trained dogs?
Observational (n=121)
Single-blind
Yes
Does a veterinary digital electrocardiograph produce different ECG values and reference intervals compared to a human digital electrocardiograph in endurance-trained dogs?
Effect estimate: Median difference 0.46 mV (95% CI 0.41 to 0.51)
Absolute Event Rate: 3.48% vs 3%
p-value: p=<0.001
Veterinary and human digital electrocardiographs produce clinically significant differences in canine ECG measurements, primarily due to differences in high-frequency filtering characteristics.
May warrant device-specific canine ECG norms; leaves open prospective validation before changing ventricular enlargement criteria.
Objective: To compare measured ECG intervals and amplitudes obtained using electrocardiographs adapted for veterinary or human use and to develop ECG reference intervals for endurance-trained dogs. Methods: Standard limb lead ECG recordings were obtained simultaneously in February and early March of 2021 using 2 digital electrocardiographs and a convenience sample of 121 endurance-trained sled adult dogs. Visually verified computer-based ECG measurements for the 2 electrocardiographs were compared using the Wilcoxon signed-rank test, Deming regression, and Bland-Altman plots. We assigned acceptable median values for differences (effect sizes) of < 10 milliseconds for P-wave duration, PQ interval, QRS duration, JT interval, and QT interval and < 10% for P-, Q-, R-, S-, and T-wave amplitudes. Results: Statistically significant differences between the 2 electrocardiographs existed for the majority (7 of 9) of global ECG indices and the majority (10 of 15) of lead I-, II-, and III-wave amplitudes. Clinically important differences based on assigned effect sizes were identified for lead II and III Q- and R-wave amplitudes, which the veterinary electrocardiograph measured 11% to 23% higher. The veterinary electrocardiograph identified a higher proportion of dogs that may have left or right ventricular enlargement. Reference intervals developed for 7 of 9 global ECG indices were narrower for the veterinary electrocardiograph. Conclusions: The low-pass filter on the veterinary electrocardiograph attenuates higher frequencies on the canine ECG to a lower extent than the human electrocardiograph. Clinical Relevance: Reference intervals for dog ECGs depend on the digital electrocardiograph, its high-frequency filtering characteristics, and its computer-based analysis program.
No takes yet. Share an insight, caveat, or question.
Constable et al. (2026) conducted an observational in Healthy endurance-trained sled dogs (n=121). Veterinary digital electrocardiograph (VET-ECG) vs. Human digital electrocardiograph (Pagewriter XLi) was evaluated on Lead II R-wave amplitude (mV) (Median difference 0.46 mV, 95% CI 0.41 to 0.51, p=<0.001). The veterinary digital electrocardiograph measured lead II and III Q- and R-wave amplitudes 11% to 23% higher than the human electrocardiograph, identifying more dogs with potential ventricular enlargement.