Key result
Ultrafast 3-D CMR proves non-inferior to standard 2-D CMR for LVEF, cutting scan time ~90%.
Why the study?
Long acquisition times and multiple breath-holds pose significant challenges in pediatric CMR, motivating evaluation of a novel 3-D ultrafast protocol against the conventional 2-D gold standard.
Does a novel 3-D ultrafast CMR protocol provide non-inferior assessment of cardiac function and strain compared to the 2-D gold standard in adolescents?
Cross-Sectional (n=39)
Single-blind
No
Does a novel 3-D ultrafast CMR protocol provide non-inferior assessment of cardiac function and strain compared to the 2-D gold standard in adolescents?
Effect estimate: Mean difference -1.2% (95% CI -2.69 to 0.29)
A novel 3-D ultrafast CMR protocol provides non-inferior quantitative assessment of cardiac function and strain in adolescents compared to the 2-D gold standard, while reducing scan time by approximately 90%.
May support abbreviated CMR in adolescents; leaves open prospective validation before practice change.
BACKGROUND: Cardiac magnetic resonance (CMR) is the gold standard for assessing cardiac anatomy and function. However, long acquisition times and multiple breath-holds pose significant challenges in pediatric imaging. OBJECTIVE: To perform a non-inferiority analysis comparing a novel 3-dimensional (D) ultrafast CMR protocol against the conventional 2-dimensional (D) gold standard for evaluating cardiac function, strain, and tissue characterization in adolescents. MATERIALS AND METHODS: Thirty-nine adolescents (mean age 12.2±2.6 years) underwent both a standard 2-D protocol and a 3-D ultrafast protocol at 3-T. The 3-D protocol comprised enhanced sensitivity encoding by static outer volume subtraction (ESSOS) cine and 3-D late gadolinium enhancement. Image quality and diagnostic confidence were compared. A pre-specified non-inferiority margin (Δ) was used to assess functional and strain parameters. Agreement was evaluated using Bland-Altman analysis and intraclass correlation coefficients (ICCs). RESULTS: The total scan time was significantly shorter for the 3-D protocol compared to the 2-D protocol (75.8±10.0 s vs. 734.4±19.6 s, P<0.05). Image quality scores were comparable between protocols (median score 4.0, P=0.74). The 3-D protocol demonstrated statistical non-inferiority for all functional and strain metrics. Bland-Altman analysis showed minimal bias for key parameters, and the 95% confidence intervals for differences met the pre-specified non-inferiority margins. Intraclass correlation coefficients (ICCs) indicated good to excellent agreement for all parameters (n=39). CONCLUSION: The novel 3-D ultrafast CMR protocol is non-inferior to the conventional 2-D gold standard for quantitative assessment of cardiac function and strain in adolescents. It offers comparable image quality with significantly reduced acquisition times, potentially improving clinical feasibility in pediatric populations.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2026) conducted a cross-sectional in Cardiovascular disease requiring cardiac magnetic resonance (n=39). 3-D ultrafast CMR protocol (ESSOS + 3-D late gadolinium enhancement) vs. Conventional 2-D CMR protocol was evaluated on Left ventricular ejection fraction (LVEF) non-inferiority (Mean difference -1.2%, 95% CI -2.69 to 0.29). The 3-D ultrafast CMR protocol was non-inferior to the 2-D gold standard for left ventricular ejection fraction (mean difference -1.2%) and reduced total scan time from 734.4 s to 75.8 s.