Key result
QardioArm blood pressure monitor meets ESH-IP2 validation requirements in patients with type 2 diabetes.
Why the study?
Home blood pressure monitoring benefits populations prone to high blood pressure, but the QardioArm mobile device had not been validated in individuals with noninsulin-dependent type 2 diabetes under European Society of Hypertension guidelines.
Does the QardioArm device accurately measure blood pressure and heart rate compared to a standard digital sphygmomanometer in patients with noninsulin-dependent type 2 diabetes?
Observational (n=33)
No
Does the QardioArm device accurately measure blood pressure and heart rate compared to a standard digital sphygmomanometer in patients with noninsulin-dependent type 2 diabetes?
The QardioArm mobile blood pressure monitoring device is valid and accurate for home monitoring of blood pressure and heart rate in patients with noninsulin-dependent type 2 diabetes.
Answers are grounded in this paper and cite their sources.
Supports QardioArm for home BP monitoring in type 2 diabetes; leaves open need for outcome trials.
BACKGROUND: Home blood pressure monitoring has many benefits, even more so, in populations prone to high blood pressure, such as persons with diabetes. OBJECTIVE: The purpose of this research was to validate the QardioArm mobile device in a sample of individuals with noninsulin-dependent type 2 diabetes in accordance with the guidelines of the second International Protocol of the European Society of Hypertension. METHODS: The sample consisted of 33 patients with type 2 diabetes. To evaluate the validity of QardioArm by comparing its data with that obtained with a digital sphygmomanometer (Omron M3 Intellisense), two nurses collected diastolic blood pressure, systolic blood pressure, and heart rate with both devices. RESULTS: The analysis indicated that the test device QardioArm met all the validation requirements using a sample population with type 2 diabetes. CONCLUSIONS: This paper reports the first validation of QardioArm in a population of individuals with noninsulin-dependent type 2 diabetes. QardioArm for home monitoring of blood pressure and heart rate met the requirements of the second International Protocol of the European Society of Hypertension.
No takes yet. Share an insight, caveat, or question.
Mazoteras‐Pardo et al. (2020) conducted an observational in Noninsulin-dependent type 2 diabetes (n=33). QardioArm blood pressure monitoring device vs. Omron M3 Intellisense was evaluated on Device validation according to the European Society of Hypertension International Protocol 2 (ESH-IP2). The QardioArm blood pressure monitoring device successfully met all validation requirements of the European Society of Hypertension International Protocol 2 in patients with type 2 diabetes.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: