Key result
Pulmonary vein isolation with the Gold-MEA catheter resulted in 60% arrhythmia-free survival at one year, with success significantly associated with a higher number of transmural lesions.
Why the study?
Does the number of transmural lesions during pulmonary vein isolation with the Gold-MEA catheter improve arrhythmia-free survival in patients with atrial fibrillation?
Cohort (n=50)
No
Does the number of transmural lesions during pulmonary vein isolation with the Gold-MEA catheter improve arrhythmia-free survival in patients with atrial fibrillation?
Answers are grounded in this paper and cite their sources.
A higher number of transmural lesions during pulmonary vein isolation with the Gold-MEA catheter is associated with improved 1-year arrhythmia-free survival in patients with atrial fibrillation.
INTRODUCTION: Pulmonary vein (PV) reconnection is frequently the cause of recurrence of atrial fibrillation (AF) after ablation. The second-generation gold multi-electrode ablation (Gold-MEA) catheter has a new design possibly resulting in improved lesion formation compared with its predecessor. We aimed to determine the association between effective radiofrequency applications with the Gold-MEA catheter and outcome after AF ablation. METHODS: , Medtronic Inc.) ablation. The Gold-MEA catheter was navigated to the PV ostium by fluoroscopy. Duty-cycled radiofrequency ablations were performed at all PV ostia. Lesions were considered transmural when electrode temperature was >50 °C and power >3 W for >30 seconds. After the ablation procedure, patients visited the outpatient clinic at 3‑month intervals including 24-hour Holter ECGs. RESULTS: Mean age was 56 years. All PVs were acutely isolated with the Gold-MEA catheter. Procedure time was 111 ± 22 minutes, ablation time was 24 ± 6.7 minutes and fluoroscopy time was 20 ± 8.1 minutes. No procedure-related complications were observed. One year after ablation, 60 % of patients were still free of arrhythmia recurrences after a single PV isolation attempt. The number of transmural lesions was associated with arrhythmia-free survival: 25.0 % in <72 transmural lesions, 64.3 % in 72-108 transmural lesions and 71.4 % in >108 transmural lesions (p = 0.029). CONCLUSION: PV isolation can be performed successfully with the Gold-MEA catheter, with a favourable safety profile. Transmurality of lesions was associated with ablation success and may improve AF ablation success.
No takes yet. Share an insight, caveat, or question.
Gal et al. (2016) conducted a cohort in Paroxysmal atrial fibrillation (n=50). Gold multi-electrode ablation (Gold-MEA) catheter was evaluated on Arrhythmia-free survival at 1 year (p=0.029). Pulmonary vein isolation with the Gold-MEA catheter resulted in 60% arrhythmia-free survival at one year, with success significantly associated with a higher number of transmural lesions.
Synapse has enriched 10 related papers. Consider them for comparative context: