Key result
Magnetocardiographic recordings detected significantly higher QRS-duration microvariability in patients with myocardial infarction compared to normals, with the highest levels in VT patients.
Why the study?
Does magnetocardiographic recording detect differences in QRS-duration microvariability among normal subjects and patients with CHD, MI, or VT?
Cross-Sectional (n=50)
Does magnetocardiographic recording detect differences in QRS-duration microvariability among normal subjects and patients with CHD, MI, or VT?
Answers are grounded in this paper and cite their sources.
Magnetocardiography can detect increased QRS microvariability in patients with MI and VT compared to normal subjects, suggesting its potential utility in identifying arrhythmogenic substrates.
High resolution electrocardiography (HRECG) recordings have already shown an increased beat-to-beat microvariability of the QRS duration of the terminal QRS in patients with a history of ventricular tachycardia (VT). The purpose of this study is to detect QRS-duration microvariability with magnetocardiographic (MCG) recordings in normals, patients with coronary heart disease (CHD), patients with a history of myocardial infarction (MI), and VT patients. QRS microvariability is calculated as the variance of time-shifts of single beats respectively to the average of all beats. The average over all channels of the MCG is performed. QRS microvariability was evaluated from 55-channel MCG in 15 normal persons, in 12 patients with CHD, in 13 patients with MI, and in 10 patients with VT. We found a significantly higher microvariability in patients with MI compared to normals. The highest microvariability was found in VT patients.
No takes yet. Share an insight, caveat, or question.
Schless et al. (2003) conducted a cross-sectional in Coronary heart disease, myocardial infarction, ventricular tachycardia (n=50). Magnetocardiographic (MCG) recordings vs. Normals was evaluated on QRS-duration microvariability. Magnetocardiographic recordings detected significantly higher QRS-duration microvariability in patients with myocardial infarction compared to normals, with the highest levels in VT patients.