Answers are grounded in this paper and cite their sources.
This article compares the performance of Fourier Domain OCT and Time Domain OCT, highlighting sensitivity advantages in optical imaging systems.
In this article we present a detailed discussion of noise sources in Fourier Domain Optical Coherence Tomography (FDOCT) setups. The performance of FDOCT with charge coupled device (CCD) cameras is compared to current standard time domain OCT systems. We describe how to measure sensitivity in the case of FDOCT and confirm the theoretically obtained values. It is shown that FDOCT systems have a large sensitivity advantage and allow for sensitivities well above 80dB, even in situations with low light levels and high speed detection.
No takes yet. Share an insight, caveat, or question.
Leitgeb et al. (2003) studied this question.