Key result
TUAS and DBS sequences showed comparable precision for in vivo 2D cardiac strain estimation, with axial strain precision requiring motion estimation rates above 544 Hz.
High motion estimation rates (>544 Hz) are necessary to optimize axial strain estimation precision in 2D cardiac ultrasonic strain imaging.
Motion estimation rates >544 Hz may optimize axial strain precision; leaves open clinical adoption of TUAS/DBS sequences.
Ultrasonic strain imaging has been applied to echocardiography and carries great potential to be used as a tool in the clinical setting. Two-dimensional (2D) strain estimation may be useful when studying the heart due to the complex, 3D deformation of the cardiac tissue. Increasing the framerate used for motion estimation, i.e. motion estimation rate (MER), has been shown to improve the precision of the strain estimation, although maintaining the spatial resolution necessary to view the entire heart structure in a single heartbeat remains challenging at high MERs. Two previously developed methods, the temporally unequispaced acquisition sequence (TUAS) and the diverging beam sequence (DBS), have been used in the past to successfully estimate in vivo axial strain at high MERs without compromising spatial resolution. In this study, a stochastic assessment of 2D strain estimation precision is performed in vivo for both sequences at varying MERs (65, 272, 544, 815 Hz for TUAS; 250, 500, 1000, 2000 Hz for DBS). 2D incremental strains were estimated during left ventricular contraction in five healthy volunteers using a normalized cross-correlation function and a least-squares strain estimator. Both sequences were shown capable of estimating 2D incremental strains in vivo. The conditional expected value of the elastographic signal-to-noise ratio (E(SNRe|ε)) was used to compare strain estimation precision of both sequences at multiple MERs over a wide range of clinical strain values. The results here indicate that axial strain estimation precision is much more dependent on MER than lateral strain estimation, while lateral estimation is more affected by strain magnitude. MER should be increased at least above 544 Hz to avoid suboptimal axial strain estimation. Radial and circumferential strain estimations were influenced by the axial and lateral strain in different ways. Furthermore, the TUAS and DBS were found to be of comparable precision at similar MERs.
No takes yet. Share an insight, caveat, or question.
Bunting et al. (2014) studied Healthy volunteers (n=5). Temporally unequispaced acquisition sequence (TUAS) and diverging beam sequence (DBS) was evaluated on 2D incremental strain estimation precision (elastographic signal-to-noise ratio). TUAS and DBS sequences showed comparable precision for in vivo 2D cardiac strain estimation, with axial strain precision requiring motion estimation rates above 544 Hz.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: