Key result
Noninvasive carotid pulse pressure evaluation using a MEMS accelerometer strongly correlated with tonometric (R=0.94, p<0.0001) and ultrasound assessments (R=0.80, p<0.0001).
Why the study?
Does an accelerometric-based approach accurately measure carotid pulse pressure compared to tonometry and ultrasound in human subjects?
Cross-Sectional (n=22)
Does an accelerometric-based approach accurately measure carotid pulse pressure compared to tonometry and ultrasound in human subjects?
Effect estimate: R = 0.94
Absolute Event Rate: 45.9% vs 47.5%
p-value: p=< 0.0001
A novel MEMS accelerometer-based method provides a valid and reproducible alternative to tonometry and ultrasound for the non-invasive assessment of carotid pulse pressure.
May support MEMS-based carotid pulse pressure monitoring; leaves open prospective validation before clinical use.
OBJECTIVE: Central pulse pressure (cPP) is increasingly investigated as possible independent predictor of cardiovascular risk and carotid pulse pressure (carPP) can be used as a surrogate of cPP. Despite its importance, carPP measurement remains challenging in clinical practice. The aim of this study was to introduce a new easier-to-use method for noninvasive carPP evaluation based on the use of a MEMS accelerometer. METHODS: carPP values (carPP(acc)) were obtained in 22 subjects (10 males, 47 ±17 years, hypertension: 50%) postprocessing and double integrating the accelerometric signals. carPP(acc) measurements were compared with tonometric assessments (carPP(ton)), and ultrasound-derived measurements (carPP(us)). Moreover, accelerometric carotid pressure waveforms (P(acc)) were contrasted in terms of shape to those obtained by tonometry (P(ton)) and ultrasound images elaboration (P(US)), calculating the root mean square error (RMSE(ton), RMSE(US)) and the regression coefficients (r(ton) and r(US)). Moreover, both the repeatability and reproducibility analyses were performed. RESULTS: carPP(acc) values (45.9 ±10.6 mmHg) were significantly correlated with carPP(ton) (47.5 ±11.3 mmHg) and carPP(US) (43.3 ±8.4 mmHg) assessments (R = 0.94, p < 0.0001 and R = 0.80, p < 0.0001, respectively). The validity of the accelerometric approach was confirmed by morphological parameters ( RMSE(ton) = 5 ±1.95 mmHg, RMSE(US) = 5.5 ±2.3 mmHg, r(ton) = 0.94 ±0.04, r(US) = 0.93 ±0.04). Coefficient of variation (CV) was equal to 6.2% for the repeatability analysis, while CV values for interoperator and intersession reproducibilities were 8.9% and 9.4%, respectively. CONCLUSION: The proposed approach, providing an easier and more available measurement, could represent a valid alternative to existing and used technique for carPP assessment.
No takes yet. Share an insight, caveat, or question.
Lascio et al. (2015) conducted a cross-sectional in Hypertension (n=22). MEMS accelerometer for carotid pulse pressure evaluation vs. Tonometry and ultrasound was evaluated on Correlation of carotid pulse pressure values between accelerometric and tonometric assessments (R = 0.94, p=< 0.0001). Noninvasive carotid pulse pressure evaluation using a MEMS accelerometer strongly correlated with tonometric (R=0.94, p<0.0001) and ultrasound assessments (R=0.80, p<0.0001).