Key result
The flexible-tip Cool Flex catheter achieved enhanced cooling (31.1 vs 36.3°C, P=0.0001) and larger ablation lesions (194.7 vs 170.9 mm3, P=0.03) compared to a rigid-tip catheter in a pig model.
Why the study?
Does a flexible-tip irrigated ablation catheter improve cooling and increase lesion size compared to a standard rigid-tip catheter in a porcine thigh muscle model?
Does a flexible-tip irrigated ablation catheter improve cooling and increase lesion size compared to a standard rigid-tip catheter in a porcine thigh muscle model?
In a preclinical porcine model, a flexible-tip irrigated ablation catheter provided better cooling, allowed more energy delivery, and created larger lesions with less charring compared to a standard rigid-tip catheter.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for flexible-tip irrigated catheters; leaves open translation to human ablation outcomes.
BACKGROUND: The flexible-tip irrigated ablation catheter Cool Flex™ (St. Jude Medical, St. Paul, MN, USA) was introduced to enhance cooling of the catheter-tissue interface and to conform to endocardial surface with better contact. Little is known about the performance of such catheter design compared to the widely used rigid-tip catheters. METHODS: In a thigh muscle preparation, ablation using the flexible-tip and rigid-tip catheters was performed in seven pigs across a range of ablation settings and catheter orientation. Postprocedure, the thigh muscle was stained with 2,3,5-triphenyltetrazolium-chloride injected into the femoral artery. The muscle was excised, fixed with formalin, and examined grossly. RESULTS: A total of 196 lesions (95 flexible tip, 101 rigid tip) were evaluated. The flexible-tip catheter was associated with enhanced cooling of catheter-tissue interface (31.1 ± 3.3°C vs 36.3 ± 3.7°C, P = 0.0001) in both perpendicular and nonperpendicular catheter orientations. This allowed more energy delivery (37.3 ± 8.9 W vs 33.7 ± 8.1 W, P = 0.004) to targeted tissue and resulted in larger lesions (median 194.7 [interquartile range: 113.1-333.8] mm(3) vs 170.9 [88.7-261.6] mm(3) , P = 0.03) than the rigid-tip catheter with larger maximum diameter (11.1 ± 2.6 mm vs 10.3 ± 2.1 mm, P = 0.03) and larger diameter at tissue surface (10.3 ± 2.4 mm vs 9.6 ± 1.7 mm, P = 0.01). Catheter orientation during ablation affected the efficiency of rigid-tip but not the flexible-tip catheter. The use of the flexible-tip catheter was associated with significantly less char formation on tissue (none vs 5.1% with rigid tip, P = 0.009). CONCLUSION: The Cool Flex™ catheter performed better than a rigid-tip catheter with enhanced cooling, larger ablation lesions, and no charring of targeted tissue.
No takes yet. Share an insight, caveat, or question.
Hussein et al. (2015) studied this question. Flexible-tip irrigated ablation catheter (Cool Flex) vs. Standard-irrigation rigid-tip catheter was evaluated on Catheter-tissue interface temperature (p=0.0001). The flexible-tip Cool Flex catheter achieved enhanced cooling (31.1 vs 36.3°C, P=0.0001) and larger ablation lesions (194.7 vs 170.9 mm3, P=0.03) compared to a rigid-tip catheter in a pig model.
Synapse has enriched 10 related papers. Consider them for comparative context: