Key result
Endocardial ablation using the FlexAbility catheter resulted in a lower risk of steam pops compared to ThermoCool SF (P=0.013) with no significant difference in lesion sizes.
Why the study?
Does the FlexAbility ablation catheter improve safety or lesion size compared to ThermoCool and ThermoCool SF catheters in a canine model?
RCT (n=12)
Chamber-level
Does the FlexAbility ablation catheter improve safety or lesion size compared to ThermoCool and ThermoCool SF catheters in a canine model?
The FlexAbility catheter produces similar lesion sizes to standard solid-tip catheters but with a lower risk of steam pops in a canine model.
Answers are grounded in this paper and cite their sources.
Animal findings do not support clinical adoption; hypothesis-generating and leaves open translation to human ablation procedures.
INTRODUCTION: Next-generation catheters have been developed to reduce irrigation volume and preserve power delivery. A novel design uses a flexible tip (FlexAbility™ catheter) that directs flow to the contact surface. Because of recent safety issues with new catheters, we undertook a study in a canine heart with 3 irrigated catheters to compare efficacy and safety. METHODS: Endocardial ablation was performed by 2 independent operators in 12 anesthetized canines with the FlexAbility (St. Jude Medical), ThermoCool™ (Biosense Webster), and ThermoCool™ SF (Biosense Webster) catheters. Endocardial RF lesions were delivered with each catheter in all 4 chambers of each animal for 52 ± 16 seconds. Each chamber was randomized to receive ablation from one catheter with recording of safety events. Cardiac pathology was performed with triphenyl tetrazolium chloride stain. RESULTS: Average lesion dimensions were not significantly different between the 3 catheters. FlexAbility™ demonstrated a lower risk of steam pops relative to ThermoCool SF (P-value = 0.013) despite equal mean power and radiofrequency time. High-temperature generator shutdowns were observed with FlexAbility™ but not with either ThermoCool catheter. High-temperature shutdowns were associated with larger average impedance drops (28.5 ohms vs. 19 ohms) without compromising lesion size. CONCLUSIONS: The FlexAbility™ tip is safe and effective with no significant difference in lesion sizes compared to both standard ThermoCool and ThermoCool SF. FlexAbility™ has a significantly lower risk of steam pops compared to ThermoCool SF in a beating heart as defined predominantly by an abrupt rise of impedance.
No takes yet. Share an insight, caveat, or question.
Winterfield et al. (2015) conducted an RCT in Endocardial ablation (n=12). FlexAbility catheter vs. ThermoCool and ThermoCool SF catheters was evaluated on Lesion dimensions and steam pops (p=0.013). Endocardial ablation using the FlexAbility catheter resulted in a lower risk of steam pops compared to ThermoCool SF (P=0.013) with no significant difference in lesion sizes.