Key result
The ThermoCool SmartTouch-SF catheter was equally effective in lesion formation compared with the SmartTouch catheter, with no significant differences in lesion volume, but provoked fewer steam pops.
Why the study?
The study was performed to compare efficacy and safety during radiofrequency ablation using ThermoCool SmartTouch and ThermoCool SmartTouch-SF catheters in the porcine heart.
Does the ThermoCool SmartTouch-SF catheter improve lesion formation and safety compared to the ThermoCool SmartTouch catheter in an ex vivo porcine heart model?
Does the ThermoCool SmartTouch-SF catheter improve lesion formation and safety compared to the ThermoCool SmartTouch catheter in an ex vivo porcine heart model?
The STSF catheter provides comparable lesion formation with a better safety profile (fewer steam pops) than the ST catheter in an ex vivo porcine model.
May favor SmartTouch-SF to limit steam pops in ablation; leaves open clinical outcome validation.
BACKGROUND: The study was performed to compare the efficacy and safety during radiofrequency ablation (RFA) using ThermoCool SmartTouch (ST) and ThermoCool SmartTouch-SF (STSF) catheters in the porcine heart. METHODS AND RESULTS: RFA was performed on the porcine myocardium by using two irrigated ablation catheters. Three groups were divided based on the different contact forces (CFs): low contact force (LCF) (1-3 g), medium contact force (MCF) (5-10 g), and high contact force (HCF) (15-20 g). In each group, RFA was delivered at four power settings of 30, 40, 50, 60 W. At each power, RFA was applied to reach the target ablation index (AI) of 350, 450, and 500. Altogether, 360 RF lesions were created by using 72 ablation conditions. AI value was positively correlated with lesion size using ST and STSF catheters. At a fixed power, lesion dimensions significantly smaller in the LCF group, whereas did not differ between MCF and HCF groups. Furthermore, at a fixed CF, lesion dimensions increased with power set at 40 W compared with 30 W but decreased with high-power RF energy (50 and 60 W). Although the average lesion surface diameter and the maximum diameter was increased using the STSF catheter, there were no significant differences in LV between the two catheters. The steam pop provoked more frequently using ST catheter and showed a negative correlation with CF and positive correlation with high-power energy. CONCLUSION: The STSF catheter is safer and equally effective in lesion formation compared with the ST catheter. LV was increased along with the early increase of CF and power, whereas a further increase of CF and power significantly reduces the lesion size.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2021) studied Radiofrequency ablation (n=360). ThermoCool SmartTouch-SF (STSF) catheter vs. ThermoCool SmartTouch (ST) catheter was evaluated on Lesion volume and steam pop frequency. The ThermoCool SmartTouch-SF catheter was equally effective in lesion formation compared with the SmartTouch catheter, with no significant differences in lesion volume, but provoked fewer steam pops.