Focal PEF ablation near metallic implants unexpectedly heats catheters up to 68°C without impedance changes.
Why the study?
The energy dynamics of pulsed electric field ablation near metallic cardiac implants during clinical procedures remain poorly understood.
Does focal PEF ablation using a lattice-tip catheter near metallic implants cause catheter overheating?
Case Report (n=3)
Does focal PEF ablation using a lattice-tip catheter near metallic implants cause catheter overheating?
Focal PEF ablation using a lattice-tip catheter near metallic implants can cause unexpected and significant catheter overheating, necessitating careful real-time temperature monitoring.
Answers are grounded in this paper and cite their sources.
Warrants real-time catheter temperature monitoring near metallic implants; hypothesis-generating for prospective PEF safety studies.
The energy dynamics of pulsed electric field (PEF) ablation near metallic cardiac implants during clinical procedures remains poorly understood. We report 3 cases in which focal PEF ablation using a lattice-tip catheter (Sphere-9) was performed adjacent to an Amplatzer septal occluder, an AtriClip, and a Sapien valve. All cases showed unexpected elevations in catheter temperature (up to 61 °C, 68 °C, and 53 °C), without impedance change or excessive bubble formation on intracardiac echocardiography. Careful real-time monitoring of catheter temperature is warranted, and intracardiac echocardiography may provide additional information when performing PEF energy delivery near metallic devices.
No takes yet. Share an insight, caveat, or question.
Hayasaka et al. (2026) conducted a case report in Patients undergoing pulsed electric field ablation near metallic cardiac implants (n=3). Focal pulsed electric field (PEF) ablation using a lattice-tip catheter (Sphere-9) was evaluated on Catheter temperature elevation. Focal pulsed electric field ablation using a lattice-tip catheter near metallic cardiac implants resulted in unexpected catheter temperature elevations up to 68 °C without impedance changes.
Synapse has enriched 10 related papers. Consider them for comparative context: