Key result
Peak PFA frequency above 261 Hz is linked to potential disappearance following initial energy delivery.
Why the study?
The optimal protocol for pulsed field ablation in AF remains unelucidated, and variables such as catheter contact require consideration to conduct safer and more effective ablation.
Does higher peak frequency at the catheter tip electrode predict effective potential disappearance during pulsed field ablation in patients with paroxysmal atrial fibrillation?
Observational (n=12)
No
Does higher peak frequency at the catheter tip electrode predict effective potential disappearance during pulsed field ablation in patients with paroxysmal atrial fibrillation?
Higher peak frequency at the catheter tip electrode during pulsed field ablation is associated with acute procedural success, suggesting its potential utility as an index for effective ablation.
Answers are grounded in this paper and cite their sources.
May inform real-time lesion assessment in PFA; leaves open validation for durable isolation or outcomes.
INTRODUCTION: Catheter ablation technology for atrial fibrillation (AF) has been improving annually, and pulsed field ablation (PFA) has recently become possible. The optimal protocol has yet to be elucidated, and considering variables, including catheter contact force, is essential for conducting more effective and safer ablation. This study was conducted using peak frequency (PF) mapping to analyze the catheter tip electrode frequency during PFA under the hypothesis that higher PF shows stronger myocardial contact and yields a more pronounced electrical current effect. METHODS: Between December 2024 and April 2025, 12 consecutive patients who underwent PFA for paroxysmal AF using the FARAPULSE system at our hospital were included in this study. Overall, 180 applications were analyzed. The PF before energy delivery was analyzed in the group in which the tip electrode potential disappeared following the first energy delivery using the flower-type configuration. RESULTS: The PF of the distal electrode in the group with potential disappearance following the first delivery was significantly higher than it was in the group without disappearance (291 ± 88 Hz vs. 267 ± 70 Hz, p = 0.02). When the potential at the tip electrode ranged from 0.5 to 1.5 mV, the area under the curve was 0.73. At a cutoff value of 261 Hz for peak frequency, sensitivity reached 67%, and specificity reached 60%. CONCLUSION: Effective PFA may be achieved using the frequency of the catheter tip electrode as an index. Delivering energy to myocardial regions with higher PF may yield greater ablation efficacy.
No takes yet. Share an insight, caveat, or question.
Aizawa et al. (2025) conducted an observational in Paroxysmal atrial fibrillation (n=12). Pulsed field ablation with peak frequency mapping vs. Applications without potential disappearance was evaluated on Peak frequency of the distal electrode associated with potential disappearance following the first energy delivery (p=0.02). Higher peak frequency during pulsed field ablation was associated with potential disappearance following the first energy delivery, with an optimal cutoff of 261 Hz.