Key result
Ferumoxytol-enhanced free-breathing 3D cine MRI matches conventional 2D breath-hold cine for measuring ventricular function.
Why the study?
Standard 2D cine CMR requires multiple breath holds that can be challenging for pediatric or ill patients, while single breath-hold alternatives yield low-resolution images that compromise cardiac function assessment.
Does Ferumoxytol-enhanced 3D cardiac cine MRI with manifold-based denoising accurately measure ventricular function compared to conventional 2D breath-hold cine in pediatric patients?
Observational (n=15)
Does Ferumoxytol-enhanced 3D cardiac cine MRI with manifold-based denoising accurately measure ventricular function compared to conventional 2D breath-hold cine in pediatric patients?
Ferumoxytol-enhanced 3D cine MRI with manifold-based denoising provides accurate ventricular function assessment without requiring breath holds, making it highly suitable for pediatric patients.
The two-dimensional (2D) cine cardiovascular magnetic resonance (CMR) technique is the reference standard for assessing cardiac function. However, one challenge with 2D cine is that the acquisition time for the whole cine stack is long and requires multiple breath holds, which may not be feasible for pediatric or ill patients. Though single breath-hold multi-slice cine may address the issue, it can only acquire low-resolution images, and hence, affect the accuracy of cardiac function assessment. To address these challenges, a Ferumoxytol-enhanced, free breathing, isotropic high-resolution 3D cine technique was developed. The method produces high-contrast cine images with short acquisition times by using compressed sensing together with a manifold-based method for image denoising. This study included fifteen patients (9.1 ± 5.6 yrs.) who were referred for clinical cardiovascular magnetic resonance imaging (MRI) with Ferumoxytol contrast and were prescribed the 3D cine sequence. The data was acquired on a 1.5T scanner. Statistical analysis shows that the manifold-based denoised 3D cine can accurately measure ventricular function with no significant differences when compared to the conventional 2D breath-hold (BH) cine. The multiplanar reconstructed images of the proposed 3D cine method are visually comparable to the golden standard 2D BH cine method in terms of clarity, contrast, and anatomical precision. The proposed method eliminated the need for breath holds, reduced scan times, enabled multiplanar reconstruction within an isotropic data set, and has the potential to be used as an effective tool to access cardiovascular conditions.
No takes yet. Share an insight, caveat, or question.
Andrews et al. (2024) conducted an observational in Cardiovascular conditions requiring MRI (n=15). Ferumoxytol-enhanced 3D cardiac cine MRI with manifold-based denoising vs. Conventional 2D breath-hold cine was evaluated on Ventricular function measurement. Ferumoxytol-enhanced free-breathing 3D cardiac cine MRI accurately measured ventricular function in 15 patients with no significant differences compared to conventional 2D breath-hold cine.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: